41 |
iASSIST 2-Pod Version
Surgical Technique
Retrieve Surgical Procedures
Archives
The system controller computer has an archiving
function to retrieve previous surgical procedure logs.
1. Insert a USB key (must be FAT32 or NTFS format) in
the USB port of the system controller computer.
Note:
Password protected USB keys are not
supported by the system controller computer.
2. On the main screen of the system controller
computer, click on the “System Utilities” tab.
3. Click on the “Copy Logs” button.
Note:
The “Copy Logs” button will not be
available until a valid USB key is connected to
the system.
4. The system will copy the logs on the USB key
and will inform the user when the archiving is
successful.
Postoperative Guide
Exit the iASSIST Application
At the end of the procedure, quit the application
before starting another procedure with another pod
kit. To do so:
1. Click on the “Stop” button at the bottom right of
the screen.
2. Confirm exit of the application by clicking “YES”
on the confirmation window that opens.
Warning:
After having scanned the barcode
and powered on the pods, do not exit the
application unless the surgical procedure has
ended. When exiting the application, the pods
will be permanently deactivated and can’t be
restarted.