iASSIST 2-Pod Version
Surgical Technique
4. Loosen the blue knob on the distal part of the tibial
alignment guide.
5. Position the rod of the tibial alignment guide to an
initial starting orientation per the preset position
(L or R) on the distal part of the tibial alignment
guide.
For a left knee
procedure, the preset position is L.
For a right knee
procedure, the preset position is R.
Figure 3.12
Tibial Registration (Continued)
Freehand Positioning of the Tibial
Adjustment Mechanism
(cont.)
3. Use three 3.5 X 38 mm hex head screw to secure
the tibial adjustment mechanism to the bone.
Note:
Secure the tibial adjustment mechanism
to the bone by first securing screws in the
upper lateral hole, secondly in the lower lateral
hole, and finally in the medial hole.
Warning:
Control the speed of the power tool
or finish fixating the screws manually to avoid
stripping the cortex of the tibia. The 500 RPM
adaptor of the Zimmer Universal Power System
Surgical Instruments can be used to secure
screws.
Warning:
From registration to validation, the
instruments must remain stable and properly
fixated to the bone to ensure accuracy of the
system.