Ziehm Imaging, Inc.
Vision Installation Manual
MAN 06-0015, Rev. A
Page 40
M a i n t e n a n c e R e p o r t : Te c h n i c a l d a ta
Ge
ner
a
to
r
Power
Direct radiography:
Fluoroscopy:
-
Pulsed mode:
40–110 kV
15 mA min./20 mA max.,
1.5 mAs min./100 mAs max.
40–110 kV
0.1–15 mA at 120 V
AC
0.2 mA at 230 V
AC
Pulse width 10, 16, 20, 30 ms
1, 2, 5, 10, 15, 30 pulses/s
Digital radiography
(snapshot):
Operating
frequency:
40–110 kV
0.1 mA min./18 mA max.
40 kHz
Max. operating data
Fluoroscopy
Direct radiography:
Digital
radiography:
110 kV / 15 mA
110 kV / 18 mA
80 kV / 20 mA
110 kV / 16 mA
80 kV / 18 mA
Max. power output
Fluoroscopy:
Direct radiography:
1650 W (110 kV / 15 mA)
1980 W (110 kV / 18 mA)
Digital
radiography:
1760 W (110 kV / 16 mA)
Nominal electric power
2000 W at 100 kV / 20 mA / 0.1 s
X-ray tube
Single-focus stationary-anode tube
Focal spot nominal size
0.6 acc. to IEC
Total filtration
≥
3 mm Al, including 0.05 mm Cu
Image intensi
fier
Tube
Input screen:
Nominal sizes:
Caesium iodide
31 / 23 / (17) cm or
23 / 17 / (10) cm
Anti-scatter grid
Pb 8/40
Содержание VISION
Страница 1: ...s Ziehm VISION Installation Manual and CDRH Report...
Страница 5: ...Table of Contents Page 2 MAN 06 0015 Chapter Rev A...
Страница 9: ...ZIEHM VISION INSTALLATION MANUAL Mounting Page 4 MAN 06 0015 Rev A 2 2...
Страница 12: ...MAN 06 0015 Rev A Page 7 ZIEHM VISION INSTALLATION MANUALInstallation This page is intentionally blank...
Страница 13: ...ZIEHM VISION INSTALLATION MANUAL CDRH Report Page 8 MAN 06 0015 Rev A...
Страница 51: ...Ziehm Imaging Inc Vision Installation Manual MAN 06 0015 Rev A Page 38 Maintenance Report Mechanical...