Vision Installation Manual
MAN 06-0015, Rev. A
Page 31
M a i n t e n a n c e R e p o r t : C h e c k C o l l i m a t i o n
14. Mark the position of the sliders on the light cross. (With a fluorescent screen,
mark the edge of the radiation field).
15. Measure the deviations of A
1
, B
1
, C and D (see Figure 20.
Fig. 21
X-Ray Field Observed on Monitor
1. Calculate the length deviation.
L =
∆Α + ∆
C
2. Calculate the width deviation.
W =
∆
B +
∆
D
3. Calculate the sum of the length and width deviation
∆
A +
∆
B +
∆
C +
∆
D
4. Calculate the percent deviation of the length, width, and length combined with
width, by multiplying their deviations by 100, and dividing the results by the SID,
97 cm, or 38.2 inches. Write the results in the box below.
Image Intensifier Format
Zoom Format
If deviations are greater than 3.2%, adjust the control system or collimator align-
ment until a deviation of less than 3.2% is achieved.
Length deviation
________ cm
__________% < 2.4%
Width deviation
________ cm
__________% < 2.4%
width deviation
________ cm
__________% < 3.2%
Length deviation
________ cm
__________% < 2.4%
Width deviation
________ cm
__________% < 2.4%
width deviation
________ cm
__________% < 3.2%
Содержание VISION
Страница 1: ...s Ziehm VISION Installation Manual and CDRH Report...
Страница 5: ...Table of Contents Page 2 MAN 06 0015 Chapter Rev A...
Страница 9: ...ZIEHM VISION INSTALLATION MANUAL Mounting Page 4 MAN 06 0015 Rev A 2 2...
Страница 12: ...MAN 06 0015 Rev A Page 7 ZIEHM VISION INSTALLATION MANUALInstallation This page is intentionally blank...
Страница 13: ...ZIEHM VISION INSTALLATION MANUAL CDRH Report Page 8 MAN 06 0015 Rev A...
Страница 51: ...Ziehm Imaging Inc Vision Installation Manual MAN 06 0015 Rev A Page 38 Maintenance Report Mechanical...