Ziehm Imaging, Inc.
Vision Installation Manual
MAN 06-0015, Rev. A
Page 22
M a i n t e n a n c e R e p o r t : M a x i m u m D o s e R a t e
Maximum Dose
Rate
Federal Regulations
Federal regulations (21 CFR 1020.32 (d) (1)) establish that fluoroscopic equipment
with automatic dose rate control shall not be operable at any combination of tube
potential and current which will result in an exposure rate in excess of 10 roentgens-
per-minute at the point where the center of the useful beam enters the patient.
1. Position a dosimeter probe as shown in Figure 10.
Fig. 10
.Set-up of Maximum Dose Rate Test
2. Select full image intensifier format.
3. Fully open the collimator.
4. Cover the image intensifier input with 3 mm of lead blocker.
5. Select
Extremities
and initiate fluoroscopy with the foot switch. Write the result-
ing dose rate in the box below.
6. Select
Pelvic
and initiate fluoroscopy with the foot switch. Write the resulting
dose rate in the box below.
7. Select
Thoracic
and initiate fluoroscopy with the foot switch. Write the resulting
dose rate in the box below..
Reject if entrance dose rate is greater than 9.9 R/min.
Extremities
Entrance dose rate = _______________ R/min
Pelvic
Entrance dose rate = _______________ R/min
Thoracic
Entrance dose rate = _______________ R/min
Содержание VISION
Страница 1: ...s Ziehm VISION Installation Manual and CDRH Report...
Страница 5: ...Table of Contents Page 2 MAN 06 0015 Chapter Rev A...
Страница 9: ...ZIEHM VISION INSTALLATION MANUAL Mounting Page 4 MAN 06 0015 Rev A 2 2...
Страница 12: ...MAN 06 0015 Rev A Page 7 ZIEHM VISION INSTALLATION MANUALInstallation This page is intentionally blank...
Страница 13: ...ZIEHM VISION INSTALLATION MANUAL CDRH Report Page 8 MAN 06 0015 Rev A...
Страница 51: ...Ziehm Imaging Inc Vision Installation Manual MAN 06 0015 Rev A Page 38 Maintenance Report Mechanical...