background image

24

25

COMPLIED STANDARDS LIST

Complied Standards List

EN ISO 14971:2012 / ISO 14971:2007 Medical devices - 

Application of risk management to medical devices

EN 1041:2008 Information supplied by the manufacturer of medical 

devices

EN 60601-1:2006+A1:2013/ IEC 60601-1:2005+A1:2012 Medical 

electrical equipment - Part 1: General requirements for basic safety 

and essential performance

EN 60601-1-11:2015/ IEC 60601-1-11:2015 Medical electrical 

equipment - Part 1-11: General requirements for basic safety and 

essential performance - Collateral standard: Requirements for medical 

electrical equipment and medical electrical systems used in the home 

healthcare environment

EN 60601-1-2:2015/ IEC 60601-1-2:2014 Medical electrical 

equipment - Part 1-2: General requirements for basic safety and 

essential performance - Collateral standard: Electromagnetic 

disturbances - Requirements and tests

EN ISO 81060-1:2012 Non-invasive sphygmomanometers - Part 1: 

Requirements and test methods for non-automated measurement type

EN 1060-3:1997+A2:2009 Non-invasive sphygmomanometers - 

Part 3: Supplementary requirements for electro-mechanical blood 

pressure measuring systems

EN 1060-4:2004 Non-invasive sphygmomanometers - Part 4: Test 

procedures to determine the overall system accuracy of automated 

non-invasive sphygmomanometers

EN 60601-1-6:2010+A1:2015/IEC 60601-1-6:2010+A1:2013 

Medical electrical equipment - Part 1-6: General requirements for basic 

safety and essential performance - Collateral standard: Usability

IEC 62366-1:2015

 Medical devices - Part 1: Application of 

usability engineering to medical devices

EN 62304:2006/AC: 2008 / IEC 62304: 2006+A1:2015   Medical 

device software - Software life-cycle processes

Risk management

Labeling
User manual

General Requirements 

for Safety

Electromagnetic

compatibility

Performance

requirements

Clinical investigation

Usability

Software life-cycle 

processes

Bio-compatibility

ISO 10993-1:2009 Biological evaluation of medical devices- Part 

1: Evaluation and testing within a risk management process

ISO 10993-5:2009 Biological evaluation of medical devices - 

Part 5: Tests for in vitro cytotoxicity

ISO 10993-10:2010 Biological evaluation of medical devices - 

Part 10: Tests for irritation and skin sensitization

EN ISO 15223-1:2016 / ISO 15223-1:2016  Medical devices. 

Symbols to be used with medical device labels, labelling and 

information to be supplied. Part 1 : General requirements

IEC 80601-2-30:2009+A1:2013 Medical electrical equipment- Part 

2-30: Particular requirements for the basic safety and essential 

performance of automated non-invasive sphygmomanometers

ISO 81060-2:2013  Non-invasive sphygmomanometers - Part 2: 

Clinical validation of automated measurement type

EMC Guidance

1)This product needs special precautions regarding EMC and needs to be 

installed and put into service according to the EMC information provided,

and this unit can be affected by portable and mobile RF communications 

equipment.

2)* Do not use a mobile phone or other devices that emit electromagnetic 

fields, near the unit. This may result in incorrect operation of the unit.

3)Caution: This unit has been thoroughly tested and inspected to assure 

proper performance and operation!

4)* Caution: This machine should not be used adjacent to or stacked with 

other equipment and that if adjacent or stacked use is necessary, this 

machine should be observed to verify normal operation in the configuration in 

which it will be used.

Table 1 

Guidance and manufacturer’s declaration – electromagnetic emissions

RF emissions 
CISPR 11

Group 1 

Class B 

Class A

Complies

Compliance

Harmonic emissions 
IEC 61000-3-2

Voltage fluctuations/ 
flicker emissions  
IEC 61000-3-3  

RF emissions 
CISPR 11

Emissions test

Electromagnetic environment - guidance

The device is intended for use in the electromagnetic environment specified below. The customer or the 

user of the device should assure that it is used in such an environment.

The device uses RF energy only for its internal 

function. Therefore, its RF emissions are very low 

and are not likely to cause any interference in nearby 

electronic equipment.

The device is suitable for use in all establishments,

other than domestic and those directly connected to 

the public low-voltage power supply network that 

supplies buildings used for domestic purposes.

Содержание UAM-710

Страница 1: ...user manual carefully and thoroughtly so as to ensure the safe usage of this product and keep the manual well for further reference in case you have problems Manufactured for Zewa Inc 12960 Commerce L...

Страница 2: ...AGEMENT 15 Recall the Records Delete the Records INFORMATION FOR USER 17 Tips for measurement Maintenances ABOUT BLOOD PRESSURE 19 What are systolic pressure and diastolic pressure What is the standar...

Страница 3: ...is intended for adult indoor use only 1 The device is not suitable for use on pregnant women 2 The device is not suitable for use on patients with implanted electrical devices such as cardiac pacemake...

Страница 4: ...to an ecchymosis Please check that operation of the device does not result in prolonged impairment of patient blood circulation When measurement please avoid compression or restriction of the connect...

Страница 5: ...ed to be calibrated within two years of reliable service If you have any problems with this device such as setting up maintaining or using please contact the SERVICE PERSONNEL of Zewa Don t open or re...

Страница 6: ...y 1 Battery powered mode 6VDC 4 AAA batteries 2 AC adaptor powered mode Not included 6V 1A Please only use the recommended AC adaptor model Please unplug the adaptor to depart from the using utility p...

Страница 7: ...ing Or when the monitor is off press SET button shortly it will display the time Then hold pressing SET button to enter the mode for year setting 2 Press the MEM to change the YEAR Each press will inc...

Страница 8: ...t the cuff is at the same level as your heart Turn your palm upwards Sit upright in a chair and take 5 6 deep breaths Rest for 5 minutes before first measuring Wait at least 3 minutes between measurem...

Страница 9: ...st three records If the records are less than three groups it will display the latest record first 2 Press the MEM or SET to get the record you want The most recent record 1 is shown first Each new me...

Страница 10: ...a coffee smoking When talking or moving your fingers When you want to discharge urine To delete records follow these steps Delete Records When the monitor is in the memory recall mode hold pressing ME...

Страница 11: ...5 or there are four or more pulse intervals the difference between each interval and the average is more than the average value of 15 the irregular heartbeat symbol appears on the display when the mea...

Страница 12: ...tion includes a list of error messages and frequently asked questions for problems you may encounter with your blood pressure monitor If the products not operating as you think it should check here be...

Страница 13: ...white backlight V A 60mm 40 5mm Approx 110mm 110mm 41mm 4xAAA batteries user manual warranty book A01 About 22cm 42cm Type BF applied part WARNING No modification of this equipment is allowed Power su...

Страница 14: ...ogical evaluation of medical devices Part 5 Tests for in vitro cytotoxicity ISO 10993 10 2010 Biological evaluation of medical devices Part 10 Tests for irritation and skin sensitization EN ISO 15223...

Страница 15: ...at 1kHz 150 kHz to 80 MHz 3 Vrms 6Vrms in ISM and amateur radio bands 80 Am at 1kHz Electromagnetic environment guidance NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These gu...

Страница 16: ...r the user of the device should assure that it is used in such an environment NOTE If necessary to achieve the IMMUNITY TEST LEVEL the distance between the transmitting antenna and the ME EQUIPMENT or...

Отзывы: