EMC GUIDANCE

26

27

 

Guidance and manufacturer’s declaration – electromagnetic immunity 

Immunity test

 

±8 kV contact
±15 kV air 

±8 kV contact
±15 kV air 

 

±2 kV 

power supply lines: 

line(s) to line(s): 

±1 kV  

line(s) to earth: 

±2 kV 

0%U

T;

 0.5 cycle

At 0°, 45°, 90°, 135°,

180°,225°,270° and 315°

0%U

T

 ;  1 cycle 

and 

70%U

T

 ; 25/30 cycles

Single phase: at 0°

0% U

T

 ; 300 cycle

30 A/m

50Hz/60Hz

NOTE      U

is the a.c. mains voltage prior to application of the test level. 

Table 2 

Compliance level

Electrostatic 
discharge 

(ESD)

 

IEC 61000-4-2 

Electromagnetic 

environment - guidance

30 A/m

50Hz/60Hz

Power frequency 

(50Hz/60Hz

magnetic field 

IEC 61000-4-8 

Voltage dips, 

short interruptions

and voltage

variations on 

power supply 

input lines

IEC 61000-4-11 

Electrical fast 
transient/burst 
IEC 61000-4-4 

Surge
IEC61000-4-5

input/output lines:

±1 kV

The device is intended for use in the electromagnetic environment specified below. The customer or the 

user of the device should assure that it is used in such an environment.

IEC 60601 test level

Floors should be wood, concrete 

or ceramic tile. If floors are 

covered with synthetic material, 

the relative humidity should be at 

least 30%.

±2 kV 

power supply lines: 

100 kHz repetition

frequency

line(s) to line(s): 

±1 kV  

100 kHz repetition

frequency

Mains power quality should be 

that of a typical commercial or 

hospital environment.

Mains power quality should be 

that of a typical commercial or 

hospital environment.

Mains power quality should be 

that of a typical commercial or 

hospital environment.

Power frequency magnetic fields 

should be at levels characteristic 

of a typical location in a typical 

commercial or hospital 

environment.

0% U

T

 ; 0.5 cycle

At 0°, 45°, 90°, 135°,

180°,225°,270° and 

315°

0% U

T

 ;  1 cycle 

and 

70% U

T

 ; 25/30 cycles

Single phase: at 0°

0% U

T

 ;300 cycle

Table 3

Guidance and manufacturer’s declaration – electromagnetic immunity

Immunity test

Compliance 

level

IEC 60601 

Test level

Conducted RF

IEC 61000-4-6 

Radiated RF 

IEC 61000-4-3

10V/m, 80% Am

at 1kHz

150 kHz to

80 MHz: 

3 Vrms

6Vrms (in ISM 

and amateur 

radio bands)

80% Am at 1kHz 

Electromagnetic environment - guidance

NOTE 1 

At 80 MHz and 800 MHz, the higher frequency range applies. 

NOTE 2 

These guidelines may not apply in all situations. Electromagnetic propagation is affected by 

absorption and reflection from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless) 

telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast 

cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to 

fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field 

strength in the location in which the device is used exceeds the applicable RF compliance level 

above, the device should be observed to verify normal operation. If abnormal performance is 

observed, additional measures may be necessary, such as re-orienting or relocating the device.

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.

b

The device is intended for use in the electromagnetic environment specified below. The customer or the 

user of the device should assure that it is used in such an environment.

150 kHz to

80 MHz: 

3 Vrms

6Vrms (in ISM 

and amateur 

radio bands)

80% Am at 1kHz 

Portable and mobile RF communications 

equipment should be used no closer to any part 

of the device, including cables, than the 

recommended separation distance calculated 

from the equation appropriate for the frequency 

of the transmitter.

Recommended separation distances:

d=0.35        ;

d=1.2

10V/m, 80% Am

at 1kHz

80 MHz to 800 MHz:

d=1.2

800 MHz to 2.7 GHz:

d=2.3

where,

 P

 is the maximum 

output power rating of the 

transmitter in watts (W) 

according to the 

transmitter manufacturer, 

d

 is the recommended 

separation distance in 

meters (m).  Field 

strengths from fixed RF 

transmitters, as 

determined by an 

electromagnetic site 

survey, should be less 

than the compliance level 

in each frequency range.

Interference may occur in 

the vicinity of equipment 

marked with the following

symbol:

P

P

P

P

EMC GUIDANCE

Содержание UAM-710

Страница 1: ...user manual carefully and thoroughtly so as to ensure the safe usage of this product and keep the manual well for further reference in case you have problems Manufactured for Zewa Inc 12960 Commerce L...

Страница 2: ...AGEMENT 15 Recall the Records Delete the Records INFORMATION FOR USER 17 Tips for measurement Maintenances ABOUT BLOOD PRESSURE 19 What are systolic pressure and diastolic pressure What is the standar...

Страница 3: ...is intended for adult indoor use only 1 The device is not suitable for use on pregnant women 2 The device is not suitable for use on patients with implanted electrical devices such as cardiac pacemake...

Страница 4: ...to an ecchymosis Please check that operation of the device does not result in prolonged impairment of patient blood circulation When measurement please avoid compression or restriction of the connect...

Страница 5: ...ed to be calibrated within two years of reliable service If you have any problems with this device such as setting up maintaining or using please contact the SERVICE PERSONNEL of Zewa Don t open or re...

Страница 6: ...y 1 Battery powered mode 6VDC 4 AAA batteries 2 AC adaptor powered mode Not included 6V 1A Please only use the recommended AC adaptor model Please unplug the adaptor to depart from the using utility p...

Страница 7: ...ing Or when the monitor is off press SET button shortly it will display the time Then hold pressing SET button to enter the mode for year setting 2 Press the MEM to change the YEAR Each press will inc...

Страница 8: ...t the cuff is at the same level as your heart Turn your palm upwards Sit upright in a chair and take 5 6 deep breaths Rest for 5 minutes before first measuring Wait at least 3 minutes between measurem...

Страница 9: ...st three records If the records are less than three groups it will display the latest record first 2 Press the MEM or SET to get the record you want The most recent record 1 is shown first Each new me...

Страница 10: ...a coffee smoking When talking or moving your fingers When you want to discharge urine To delete records follow these steps Delete Records When the monitor is in the memory recall mode hold pressing ME...

Страница 11: ...5 or there are four or more pulse intervals the difference between each interval and the average is more than the average value of 15 the irregular heartbeat symbol appears on the display when the mea...

Страница 12: ...tion includes a list of error messages and frequently asked questions for problems you may encounter with your blood pressure monitor If the products not operating as you think it should check here be...

Страница 13: ...white backlight V A 60mm 40 5mm Approx 110mm 110mm 41mm 4xAAA batteries user manual warranty book A01 About 22cm 42cm Type BF applied part WARNING No modification of this equipment is allowed Power su...

Страница 14: ...ogical evaluation of medical devices Part 5 Tests for in vitro cytotoxicity ISO 10993 10 2010 Biological evaluation of medical devices Part 10 Tests for irritation and skin sensitization EN ISO 15223...

Страница 15: ...at 1kHz 150 kHz to 80 MHz 3 Vrms 6Vrms in ISM and amateur radio bands 80 Am at 1kHz Electromagnetic environment guidance NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These gu...

Страница 16: ...r the user of the device should assure that it is used in such an environment NOTE If necessary to achieve the IMMUNITY TEST LEVEL the distance between the transmitting antenna and the ME EQUIPMENT or...

Отзывы: