background image

19

18

ABOUT BLOOD PRESSURE

INFORMATION FOR USER

What are systolic pressure and diastolic pressure? 

press

artery

vein

blood discharging

Systolic

relax

blood entering

Diastolic  

When ventricles contract and pump blood out of the 

heart, the blood pressure reaches its maximum value 

in the cycle, which is called systolic pressure. When 

the ventricles relax, the blood pressure reaches its 

minimum value in the cycle, which is called diastolic 

pressure.

Irregular Heartbeat Detector

CAUTION

The appearance of the IHB icon indicates that a pulse irregularity consistent with an irregular heart-

beat was detected during measurement. Usually this is NOT a cause for concern. However, if the 

symbol appears often, we recommend you seek medical advice. Please note that the device does 

not replace a cardiac examination, but serves to detect pulse irregularities at an early stage.

An irregular heartbeat is detected when a heartbeat rhythm varies while the unit is measuring

the systolic and diastolic blood pressure.During each measurement, the monitor records all the

pulse intervals and calculate the average ; if there are two or more pulse intervals ,the difference

between each interval and the average is more than the average value of ±25% , or there are four

or more pulse intervals ,the difference between each interval and the average is more than the

average value of ±15%,the irregular heartbeat symbol appears on the display when the

measurement results are appeared.

What is the standard blood pressure classification?

The chart on the right is the 

standard blood pressure classifi-

cation published by American

Heart Association (AHA).

Please consult a physician if your measuring result falls outside the range. 

Please note that only a physician can tell whether your blood pressure value 

has reached a dangerous point.

CAUTION

Blood Pressure Category

Normal

Elevated

High Blood Pressure

(Hypertension) Stage 1

High Blood Pressure

(Hypertension) Stage 2

Hypertensive Crisis

(Consult your doctor immediately)

Systolic

mmHg (upper#)

Diastolic

mmHg (lower#)

less than 120

120-129

130-139

140 or higher

Higher than 180

and

or

or

and/or

less than 80

80-89

90 or higher

Higher than 120

This chart reflects blood pressure categories defined by American Heart Association.

and

less than 80

Maintenance

In order to get the best performance, please follow the  

instructions below.

Put in a dry place and avoid the sunshine

Avoid intense shaking

and collisions

Using wet cloths to remove dirt

Avoid touching water,

clean it with a dry cloth in case.

Avoid dusty and unstable

temperature environment

Do not attempt to clean the reusable cuff 

with water and never immerse the cuff in 

water.

Содержание UAM-710

Страница 1: ...user manual carefully and thoroughtly so as to ensure the safe usage of this product and keep the manual well for further reference in case you have problems Manufactured for Zewa Inc 12960 Commerce L...

Страница 2: ...AGEMENT 15 Recall the Records Delete the Records INFORMATION FOR USER 17 Tips for measurement Maintenances ABOUT BLOOD PRESSURE 19 What are systolic pressure and diastolic pressure What is the standar...

Страница 3: ...is intended for adult indoor use only 1 The device is not suitable for use on pregnant women 2 The device is not suitable for use on patients with implanted electrical devices such as cardiac pacemake...

Страница 4: ...to an ecchymosis Please check that operation of the device does not result in prolonged impairment of patient blood circulation When measurement please avoid compression or restriction of the connect...

Страница 5: ...ed to be calibrated within two years of reliable service If you have any problems with this device such as setting up maintaining or using please contact the SERVICE PERSONNEL of Zewa Don t open or re...

Страница 6: ...y 1 Battery powered mode 6VDC 4 AAA batteries 2 AC adaptor powered mode Not included 6V 1A Please only use the recommended AC adaptor model Please unplug the adaptor to depart from the using utility p...

Страница 7: ...ing Or when the monitor is off press SET button shortly it will display the time Then hold pressing SET button to enter the mode for year setting 2 Press the MEM to change the YEAR Each press will inc...

Страница 8: ...t the cuff is at the same level as your heart Turn your palm upwards Sit upright in a chair and take 5 6 deep breaths Rest for 5 minutes before first measuring Wait at least 3 minutes between measurem...

Страница 9: ...st three records If the records are less than three groups it will display the latest record first 2 Press the MEM or SET to get the record you want The most recent record 1 is shown first Each new me...

Страница 10: ...a coffee smoking When talking or moving your fingers When you want to discharge urine To delete records follow these steps Delete Records When the monitor is in the memory recall mode hold pressing ME...

Страница 11: ...5 or there are four or more pulse intervals the difference between each interval and the average is more than the average value of 15 the irregular heartbeat symbol appears on the display when the mea...

Страница 12: ...tion includes a list of error messages and frequently asked questions for problems you may encounter with your blood pressure monitor If the products not operating as you think it should check here be...

Страница 13: ...white backlight V A 60mm 40 5mm Approx 110mm 110mm 41mm 4xAAA batteries user manual warranty book A01 About 22cm 42cm Type BF applied part WARNING No modification of this equipment is allowed Power su...

Страница 14: ...ogical evaluation of medical devices Part 5 Tests for in vitro cytotoxicity ISO 10993 10 2010 Biological evaluation of medical devices Part 10 Tests for irritation and skin sensitization EN ISO 15223...

Страница 15: ...at 1kHz 150 kHz to 80 MHz 3 Vrms 6Vrms in ISM and amateur radio bands 80 Am at 1kHz Electromagnetic environment guidance NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These gu...

Страница 16: ...r the user of the device should assure that it is used in such an environment NOTE If necessary to achieve the IMMUNITY TEST LEVEL the distance between the transmitting antenna and the ME EQUIPMENT or...

Отзывы: