GUIDANCE AND MANUFACTURER’S DECLARATION-ELECTROMAGNETIC EMISSIONS & IMMUNITY | 36
GUIDANCE AND MANUFACTURER’S DECLARATION-ELECTROMAGNETIC
IMMUNITY
The device is intended for use in the electromagnetic environment specified below. The customer
or the user of the device should assure that the device is used in such an environment.
Immunity test
IEC60601-1-2 test level
Compliance level
Electromagnetic environment-guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 2 kV for power
supply lines
± 1 kV for
input/output lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV differential
mode
± 2 kV common mode
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% UT (>95% dip in UT)
for 0.5 cycle
40% UT (60% dip in UT)
for 5 cycles
70%
U
T (30% dip in
U
T)
for 25 cycles
<5% UT (>95% dip in UT)
for 5 sec
<5% UT (>95% dip in
UT) for 0.5 cycle
40% UT (60% dip in UT)
for 5 cycles
70%
U
T (30% dip in
U
T)
for 25 cycles
<5% UT (>95% dip in
UT) for 5 sec
Mains power quality should be that of a
typical commercial or hospital
environment.
If the user of the device requires continued
operation during power mains
interruptions, it is recommended that the
device be powered from an uninterruptible
power supply.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital
environment.
Portable and mobile RF communications
equipment should be used no closer to any
part of the device including cables, than
the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
d = 1.2
√
P
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
d = 1.2
√
P 80 MHz to 800 MHz
d =
2.3
√
P
800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W) per
the transmitter manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey, ª should be less than the
compliance level in each frequency range.
b
Interference may occur near equipment
marked with the following symbol: