GUIDANCE AND MANUFACTURER’S DECLARATION-ELECTROMAGNETIC EMISSIONS & IMMUNITY | 36 

GUIDANCE AND MANUFACTURER’S DECLARATION-ELECTROMAGNETIC 
IMMUNITY 

The device is intended for use in the electromagnetic environment specified below.  The customer 
or the user of the device should assure that the device is used in such an environment. 
 

Immunity test 

IEC60601-1-2 test level 

Compliance level 

Electromagnetic environment-guidance 

Electrostatic 

discharge (ESD) 
IEC 61000-4-2 

± 6 kV contact 
± 8 kV air 

± 6 kV contact 
± 8 kV air 

Floors should be wood, concrete or 

ceramic tile. If floors are covered with 
synthetic material, the relative humidity 

should be at least 30 %. 

Electrical fast 

transient/burst 
IEC 61000-4-4 

± 2 kV for power supply 
lines 

± 1 kV for input/output 
lines 

± 2 kV for power 
supply lines 

± 1 kV for 
input/output lines 

Mains power quality should be that of a 

typical commercial or hospital 
environment. 

Surge 
IEC 61000-4-5 

± 1 kV line(s) to line(s) 
± 2 kV line(s) to earth 

± 1 kV differential 
mode 

± 2 kV common mode 

Mains power quality should be that of a 
typical commercial or hospital 

environment. 

Voltage dips, 

short 
interruptions and 

voltage variations 
on power supply 
input lines 

IEC 61000-4-11 

<5% UT (>95% dip in UT) 
for 0.5 cycle 

40% UT (60% dip in UT) 
for 5 cycles 

70% 

U

T (30% dip in 

U

T) 

for 25 cycles 

<5% UT (>95% dip in UT) 
for 5 sec 

<5% UT (>95% dip in 
UT) for 0.5 cycle 

40% UT (60% dip in UT) 
for 5 cycles 

70% 

U

T (30% dip in 

U

T) 

for 25 cycles 

<5% UT (>95% dip in 
UT) for 5 sec 

Mains power quality should be that of a 
typical commercial or hospital 

environment.  
If the user of the device requires continued 

operation during power mains 
interruptions, it is recommended that the 

device be powered from an uninterruptible 
power supply. 

Power frequency 
(50/60 Hz) 

magnetic field 
IEC 61000-4-8 

3 A/m 

3 A/m 

Power frequency magnetic fields should be 

at levels characteristic of a typical location 
in a typical commercial or hospital 

environment. 
 

Portable and mobile RF communications 
equipment should be used no closer to any 

part of the device including cables, than 
the recommended separation distance 

calculated from the equation applicable to 
the frequency of the transmitter. 

 

Recommended separation distance 

Conducted RF 
IEC 61000-4-6 

3 Vrms 
150 kHz to 80 MHz 

3 Vrms 

d = 1.2

Radiated RF 

IEC 61000-4-3 

3 V/m 

80 MHz to 2.5 GHz 

3 V/m 

d = 1.2

P 80 MHz to 800 MHz 

 

 

 

d = 

2.3

800 MHz to 2.5 GHz 

 
where P is the maximum output power 

rating of the transmitter in watts (W) per 
the transmitter manufacturer and d is the 

recommended separation distance in 
meters (m). 
Field strengths from fixed RF transmitters, 

as determined by an electromagnetic site 
survey, ª should be less than the 

compliance level in each frequency range.

 

b

  

Interference may occur near equipment 
marked with the following symbol:  

 

Содержание Z6

Страница 1: ...Non Prescriptive Non invasive Fat Loss Laser INSTALLATION AND PROPER USE REFERENCE GUIDE Read this entire booklet before using your Zerona Z6 OTC Laser...

Страница 2: ......

Страница 3: ...44 0 1491 821135 Email info erchoniaeurope com Or visit erchonia com ATTENTION This device is sold Over the Counter OTC In general there is no prescription required for the use of this device by any...

Страница 4: ...RECTIONS 12 BACKUP BATTERIES 15 SECTION 3 ZERONA Z6 OTC OPERATION 16 BASICS AND ESSENTIALS 16 MOVING THE LASER ARM 16 WHEEL LOCKS 16 ZERONA Z6 TOUCH SCREEN 17 DEVICE INFORMATION SCREEN 19 PROTECTIVE E...

Страница 5: ...GUIDANCE AND MANUFACTURER S DECLARATION ELECTROMAGNETIC EMISSIONS IMMUNITY 35 TECHNICAL SPECIFICATIONS 38...

Страница 6: ...uction of body circumference The Zerona Z6 OTC laser delivers a new non invasive fat loss procedure designed to slim the body without surgery pain or needles The protocol allows clients to continue da...

Страница 7: ...When you receive the shipment carefully inspect the container for damage If the shipping container or cushion material is damaged keep it until the contents have been checked for completeness and the...

Страница 8: ...lerts you about a situation which if not avoided could result in death or serious injury It may also describe potential serious adverse reactions and safety hazards Precaution is used for the statemen...

Страница 9: ...k fire or personal injury To avoid risk of electric shock do not defeat the safety purpose of the grounding type plug A grounding type plug has two blades and a third grounding prong The wide blade or...

Страница 10: ...e the specified level of protection against spilled or splashed liquids unplug the device from the power supply and thoroughly dry all exposed surfaces of this device and allow to dry thoroughly prior...

Страница 11: ...at emit frequency If this device causes interference to other devices which can be determined by turning the device off and on you are encouraged to try to correct the interference by one of the follo...

Страница 12: ...e governing agencies requires specific labeling All required labels are affixed per the relevant codes Each label is pictured and described in this section Additionally the placement of each label on...

Страница 13: ...er no pushing reminder and each of the laser diodes 6 have a label affixed that show the direction of the laser beam output The following diagram shows one of the 2 identified pinch points the device...

Страница 14: ...the laser device is that long term exposure to laser light could cause damage to eyesight As a precaution when you are giving the treatments with the Zerona Z6 OTC ensure that the client wears the sp...

Страница 15: ...rence reduction of the waist hips and thighs with 6 treatments over a 2 week time frame A double blind randomized multi site and placebo controlled study with 67 participants was conducted in which pa...

Страница 16: ...6 OTC ASSEMBLY DIAGRAM OF MACHINE Fig 1 1 Laser Head Assembly 2 Laser Output Head 3 Power Safety Lockout Key 4 Laser Arm 5 Arm Lock 6 Touch Screen 7 Main Upright of Base 8 Power Inlet Fuse Holder 9 Wh...

Страница 17: ...rom the main upright of base Once the locking knob is removed hold the laser head assembly and laser arm piece while you attach the base 7 and the arm piece 4 Fig 2 Gently feed the laser arm cable int...

Страница 18: ...ush and the hole opening for the locking knob is aligned INCORRECT CORRECT With the holes aligned thread the locking knob into place This will keep the main head assembly from unwanted rotation during...

Страница 19: ...e the cover Now that the laser has been put together insert the electrical power cord 12 into the power inlet 8 located at touch screen side of the base device See figure 1 Plug the electrical power c...

Страница 20: ...ture to ensure no unauthorized use of the laser device The device will not operate in the OFF position Turning the key to the OFF position while using the device will immediately power off the device...

Страница 21: ...facture to a neutral state This means that the arm will not move up or down unless you physically move it Once you have maneuvered the laser arm to the desired position as a secondary precaution use t...

Страница 22: ...ate icon Make sure the power is connected Turn the device ON by placing one of the keys into the key lock and turning it clockwise to the ON position When the device is ready for use after about 30 45...

Страница 23: ...e text will display PRESS TO RESUME To resume the procedure touch the flashing PRESS TO RESUME icon Zerona Z6 Mode Screen Zerona Z6 Halfway Point Screen Once the Time Remaining display reaches 20 00 t...

Страница 24: ...k without resetting the time remaining Touch the PRESS TO RESUME icon to continue with treatment Zerona Z6 Mode Screen DEVICE INFORMATION SCREEN To access the Device Information screen touch the Ercho...

Страница 25: ...y into the eye and maintaining it there for an extended period could prove to be damaging To ensure there is no possible instance of residual effect we have included a pair of specialty glasses for us...

Страница 26: ...he center diodes slightly below the navel Position the adjustable laser arms so they are targeting the greatest accumulation of fat on the left and right side of the waist hips and thighs NOTE It is i...

Страница 27: ...GENERATED BEAM PARAMETERS 3 0 Inch 7 5 cm Line Generated Beams 6 inches 15 cm separating adjacent Line Generated Beams Use the palm of your hand to position the adjustable laser heads shifting the fle...

Страница 28: ...t treatment areas of fat Position the adjustable laser arms so they are targeting the greatest accumulation of fat on the left and right side of the waist hips and thighs NOTE It is important to align...

Страница 29: ...th the lasers on check the positioning of the heads for coverage of targeted areas as described in the previous section of this booklet When the first 20 minutes of treatment on the client s front bod...

Страница 30: ...the answer as Yes or No Question Yes No Are you under 18 years of age Female clients only Are you pregnant or do you think you might be pregnant Not applicable client is male Do you have any open woun...

Страница 31: ...d the client s weight in pounds that you measured This number may fall in between two other numbers Move your finger to the closest number which may be the number above or below the client s weight nu...

Страница 32: ...ZERONA Z6 OTC OPERATION 27...

Страница 33: ...stand with their feet shoulder width apart Have them relax their muscles Do not let them suck it in as this is a typical reflex of most clients When taking measurements make sure that the tape is pull...

Страница 34: ...R SERVICE Before sending a device to the Erchonia Service and Repair Department for repair obtain a service order SO number from the service representative Pack the device in the original containers i...

Страница 35: ...s a touchscreen interface periodic cleaning of the touchscreen will be necessary To clean the touchscreen use a nearly dry cloth containing one of the mild cleaning agents listed below Ensure there is...

Страница 36: ...To achieve the specified level of protection against spilled or splashed liquids unplug the device from the power supply and thoroughly dry all exposed surfaces of this device and allow to dry thorou...

Страница 37: ...Shipping required to facilitate warranty repair and or maintenance issues within the first 90 days will be paid by manufacturer Shipping required to facilitate warranty repair and or maintenance issue...

Страница 38: ...ase contact our Erchonia Customer Care representative for assistance Contact us at Erchonia Customer Care Phone 1 321 473 1251 Email info erchonia com Or visit erchonia com Copyright 2016 Erchonia Cor...

Страница 39: ...ctly O OFF and ON Blown fuse or tripped circuit breaker in building Contact a qualified service technician Diode s does not power on Electrical connector came apart Remove head assembly from base asse...

Страница 40: ...Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The device uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any inte...

Страница 41: ...25 cycles 5 UT 95 dip in UT for 5 sec 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec Mains power quality should be that o...

Страница 42: ...ENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT AND THE DEVICE The device is intended for use in an electromagnetic environment in which radiated RF disturbances are...

Страница 43: ...Rated operating altitude 2000 m Normal barometric pressure 80 0 kPa Erchonia Corporation 650 Atlantis Rd Melbourne FL 32904 USA QNET BV Kantstraat 19 NL 5076 NP Haaren The Netherlands Trade name Zeron...

Страница 44: ...6 OTC Installation and Proper Use Reference Guide You are now ready to provide great outcomes with the Zerona Z6 OTC Non Prescriptive Non invasive Fat Loss Laser It s just that simple Learn more about...

Отзывы: