5.1 INTRODUCTION
Before the D2O can perform its normal operation it must be programmed with
the desired operating parameters.
Programming occurs via a serial interface, using a plaintext protocol that can
be entered by either human or automated system input. A Serial Terminal
software application (such as
Tera Term
), running on a Computer, is required
in order to send human entered input commands via a Command Line
Interface (CLI) to the D2O.
Once a connection is established, the CLI provides the following setup
features:
•
Control of the D2O Digital Power Switch
•
Set device acquisition and system parameters
•
Generate a Commissioning Report to confirm current settings
When setting up a D2O for the first time the steps shown in the following
sections should be sequentially followed to ensure the D2O will be correctly
setup and perform its normal operation.
Before commencing any programming or commissioning tasks the following
prerequisite tasks must be completed:
•
A
4-Pin M12 to USB-A Cable
must be available for connection between the
D2O and a USB port of the Computer, without extension. Ensure the USB
Port is capable of providing at least a 500mA supply
•
Ensure GPS satellite signals are available by moving outdoors or close to a
window (required to obtain date, time and location settings of the D2O)
•
Serial Terminal software is installed on the Computer
5.2 PRE-PROGRAMMING CHECKLIST
TO COMPUTER
5.3 CONNECTING TO A COMPUTER
Interfacing to the D2O can occur once the unit has been connected via the
4-Pin M12 to USB-A Cable,
to the Computer.
Remove the protective
Dust Cap
from the
4-Pin M12 Connector
on the
bottom of the D2O, and store safely for later replacement. Connect the
4-Pin
M12 to USB-A Cable
in-between the D2O and a Computer, ensure that the
Computer’s USB Port is capable of providing at least a 500mA supply.
The D2O unit is now being powered by the USB connection and will
commence its start-up sequence.