COOLTONE User Manual
Chapter 4: Cleaning and Maintenance
-C
35
Mechanical
cleaning of the
securement
system
Pre-Cleaning:
Turn over the applicator sleeves to be sure that they do not stick to
other clothes. Close the Velcro of the applicator sleeves and the
strap.
Cleaning:
1. Do not bleach
2. Tumble dry gentle low heat. Use very mild treatment.
3. Do not iron
4. Do not dry clean
5. Machine wash cold gentle cycle.
Reusability:
Over time, the material can become damaged depending on use
and treatment.
Once the securement system is punctured, worn, has optical
damages, has non-removable contaminations or the Velcro is not
functional anymore, please do not reuse it.
Disinfection
We recommend that disinfection be carried out before every treatment and at the end of every day as
well as in the event of evidence of possible contamination. Consult with your health professional when
doing so. Always perform cleaning prior to disinfection.
Monitor housing and applicators can be disinfected using disinfectant wipes. Use a commercially
available alcohol-free disinfectant for metal and plastic, with bactericidal, viricidal and fungicidal
properties. Observe the application instructions of the manufacturer. Wipe all surfaces using a cloth
soaked according to the specifications of the manufacturer of the disinfectant, but not dripping, or with
cloth pre-impregnated with disinfectant (wipes).
If applicable, also observe requirements for drying or post-cleaning.
Caution
If flammable solutions are used for cleaning and disinfection, sufficient time must be allowed for
the solutions to evaporate before using the device. Otherwise, it may lead to flammability.
Notice
Only use the device in a clean environment