Chapter 3: Error Messages / Troubleshooting
-C
33
Self Trouble-Shooting
If the system encounters a problem, it displays a message to help you diagnose and resolve the issue.
When an error message occurs, carry out the recommended action, if any. If the problem persists,
record both codes and contact Customer Service (see page 5).
For assistance with any message not listed here, contact Customer Service (see page 5).
Table 2: Self trouble-shooting
Symptom
Action
Equipment does not turn on.
Check if the power connector of the
equipment is properly connected.
Check if the power switch of
equipment is turned on.
Magnetic field is not generated from equipment
Confirm if the LCD displays an
ERROR message.
Check if the output strength is set
by checking the intensity values
after pressing the Start button
.
Over Temperature message is displayed.
If applicator is in an upright
position, it is easy for the applicator
to overheat. Maintain a horizontal
position as much as possible.
If the room temperature is too high,
disorder can be caused in the
cooling. Maintain the room
temperature at less than 28
℃
as
much as possible.
The main power switch spontaneously turns off.
The LCD screen of operation panel does not
illuminate when power is turned off and then
turned on again.
Stimulation is not generated by applicator, even
after intensity is increased.
The temperature icon on the screen blinks and the
equipment is not operating.
In any case of oil leakage.
Stop the operation by cutting off
the power to the device and
contact the service center.