Chapter 3: Error Messages / Troubleshooting
COOLTONE User Manual
32
-C
Table 1: Error Messages and Action
Message
number
Message text
Action
T100-YYY
Internal Temperature
Wait until the system is cooled down.
T200-YYY
Applicator Error
Power the control unit off, connect both applicators and
power it on.
T201-YYY
Control Unit Error
Remove the USB stick if inserted and power the control
unit off and on.
T230-YYY
No treatments
remaining
Connect a new card.
T231-YYY
Card error
Disconnect and reconnect the card.
T500-YYY
Chiller Error
Power the control unit off and on.
If the problem persists, contact Customer Service.
T501- YYY
Applicator Error
Power the control unit off and on.
If the problem persists, contact Customer Service.
T502- YYY
Control Unit Error
Power the control unit off and on.
If the problem persists, contact Customer Service.
T530-YYY
Control Unit Error
Power the control unit off and on.
If the problem persists, contact Customer Service.
T531-YYY
Control Unit Error
Contact customer service.
T532-YYY
Software Installation
Error
Power the control unit off and on and retry the
installation.
If the problem persists, contact Customer Service.
T533-YYY
Software Installation
Error
Software installation cannot be performed as installation
is an older version than the system.
Remove the USB stick if inserted and power the control
unit off and on.
T534-YYY
Software Installation
Error
Power the control unit off and on and retry the
installation.
If the problem persists, contact Customer Service.