2 Requirements | 2.2 Location Requirements
ZEISS
Entrance
Min. 0.8 m wide
Hallways
Min. 1.2 m wide
Corners
Min. 1.2 m for boxes
Transport ways
Free of staircases
Power sockets
Should be no further than 1.5 m from the
power supply entry socket of the Microscope
System
§
Allow for sufficient space depending on the size of the tables used on site.
§
Electrical components must be at least 10 cm away from the wall and not near flammable ob-
jects.
The following layouts are recommended for optimal accessibility, airflow and stability, depending
on the options purchased with the system. For possible changes in preferential layout, please dis-
cuss with your ZEISS representative. The placing of the user table can vary depending on available
space and cable lengths.
Service
Access Area
850
500
2350
3300
800
1000
1000
1200
400
500
5
7
4
3
2
1
max. 1500
6
Fig. 1: Recommended arrangement for Axio Imager.Z2 or Axio Examiner.Z1 (measurements in
mm)
1
LSM scan module
2
Component rack with laser module and
power supply unit
3
System table, small
4
Microscope stand
§
Axio Imager.Z2 or
§
Axio Examiner.Z1
5
Power point for accessories
6
Gas supplies (if required)
7
Customer arranged computer
workspace
8