2 Requirements | 2.2 Location Requirements

2.1.4 Vibration

Vibrations can be caused e.g. by heavy-duty machinery installed on the same floor or even in the
same building as well as transport facilities operated nearby. Depending on the floor stability and
construction, even walking in the room or in the hallways may affect the image quality.

Allowable vibration values (under operating conditions on system table):

§

Less than Vibration Class VC-C

, 12.5 μm/s (IEST RP 12 and ISO 10811)

Reference for floor measurements:

§

Please conform with the attenuation specifications of the proposed system table to ensure the
VC-C specification is met on the table.

§

For the NEWPORT Vision IsoStation series of damping tables recommended by ZEISS:
Less than 63 µm/s for 0 to 4 Hz and less than 125 µm/s for 4 Hz to 1000 Hz

§

These recommendations are based on an attenuation of -10 dB starting from 4 Hz and over
-26 dB above 10 Hz. Low frequency vibrations (swaying) is recommended to be less than half
of the 4 Hz limits.

Requirements for measuring vibrations:

§

Three positions should be measured.

§

The positions need to cover a triangle over the area where the microscope will be installed.

§

At each position, measurements in X, Y and Z need to be taken.

§

X is along the front of the microscope.

2.2 Location Requirements

The Microscope System must only be operated in closed rooms. It is recommended to install the
Microscope System in a dimmed room where artificial illumination, sunlight or other light sources
cannot interfere with image acquisition. The Microscope System should not be installed near radi-
ators or windows with direct sunlight. Compliance with the installation requirements of the Mi-
croscope System and the availability of the requested supplies is the responsibility of the customer
and has to be readily available at the time of installation.

Installation site

Exclusively inside buildings

Operating range

Max. 2000 m above sea level

§

Do not place the Microscope System near windows with direct sunlight or radiator heating
systems which could change the temperature of the system.

§

Do not position power supplies or racks under the system table because the heat dissipated
can cause instability of the Microscope System through the expansion of the steel table.

§

Harmful gases and fumes that may be used on the Microscope System must be safely ex-
tracted from the Microscope System and site in accordance with local safety regulations.

§

The Microscope System must not be set up in areas with potential danger by explosives.

2.2.1 Space Requirements

Recommended room size

Min. 4.0 m x 5.0 m x 2.3 m

Service and safety clearance area

Min. 0.5 m around the Microscope System

Min. 0.9 m clearance from the table edge on
the side of the laser module

Customer arranged computer workspace

Min. 0.8 m x 1.2 m x 0.76 m

Installation Requirements ZEISS LSM 900 | en-US | Rev. 2 | 349500-8183-000

7

Содержание LSM 900

Страница 1: ...Installation Requirements ZEISS LSM 900 Laser Scanning Microscope...

Страница 2: ...archiving purposes shall remain unaffected thereby Any viola tions may be prosecuted as copyright infringements The use of general descriptive names registered names trademarks etc in this document d...

Страница 3: ...rvoltage Category 10 2 3 2 UPS Requirements 12 2 4 Network Connectivity 12 2 4 1 ZEISS Predictive Service 13 2 5 Safety and Radiation Information 14 2 5 1 Laser Safety 14 2 5 2 Biosafety Level of the...

Страница 4: ...oom temperature Air quality The customer is responsible for ensuring that the conditions specified in this document are met and that the required equipment is available before delivery and installatio...

Страница 5: ...than 50 dB Z class 200 to 500 Hz less than 55 dB Z class 2 1 2 Air Conditioning and Quality The following nominal power usage for the Microscope System and possible accessories is shown in Watts To m...

Страница 6: ...porary conductivity caused by condensation is to be expected Example A circuit used in an of fice environment The circuitry inside a computer would fall in this cate gory 3 Conductive pollution occurs...

Страница 7: ...d in closed rooms It is recommended to install the Microscope System in a dimmed room where artificial illumination sunlight or other light sources cannot interfere with image acquisition The Microsco...

Страница 8: ...depending on the options purchased with the system For possible changes in preferential layout please dis cuss with your ZEISS representative The placing of the user table can vary depending on avail...

Страница 9: ...ds and supplied socket with protective ground contact securely connected The provided electrical connection must be in accordance with the applicable electrical codes for the country of installation I...

Страница 10: ...eferred to as Installation Category to address transient voltages Category IV devices can handle the largest transients relative to the normal working voltage Category I devices can handle only small...

Страница 11: ...oltage levels of these loads must be rated at this overvoltage category III In fixed installations and for cases where the reliability and the availability of the equipment is subject to special requi...

Страница 12: ...til normal mains is restored Please ensure to switch off the Microscope System be fore the batteries run low in a blackout In areas where the mains supply is very unstable or not very clean the most s...

Страница 13: ...re the above network connectivity and predictive service will oper ate under the parameters shown in the table below Ports 443 Protocols HTTPS secure websocket URL of ZEISS server predictive service z...

Страница 14: ...s the responsible body the operator is responsible for complying with laser safety regula tions The following types of lasers are currently intended for use Using other lasers than indicated be low re...

Страница 15: ...ever some do exist in a small number of places in the US and around the world As the highest level of biological safety a BSL 4 lab consists of work with highly dangerous and exotic mi crobes Infectio...

Страница 16: ...e user if the air supplied is from the building facilities or from gas bottles Due to acoustic noise and vibrations where possible the compressor should be installed in a sep arate room close to the w...

Страница 17: ...f the system where necessary Some components such as the tables are large heavy or bulky and may require extra assis tance to get the units into the allocated site Maximum shock resistance Do not drop...

Страница 18: ...s used not possi ble for Airyscan Anti vibration plate for Axio Imager 320 300 45 7 Anti vibration plate for Axio Observer 800 525 45 7 Microscope Axio Examiner Z1 700 390 820 24 Axio Imager Z2 560 39...

Страница 19: ...EU RoHS and 2012 19 EU WEEE as applicable for the product In addition to the European and international guidelines and standards the 21 CFR 1040 10 Performance Standards for light emitting products la...

Страница 20: ...ch requirements are yet to be completed ZEISS can be requested to assist with the checking of the environmental conditions Environmental conditions in and around a system can change with time due to v...

Страница 21: ...adiation Information X Ray Laser Safety Biosafety Level of the Laboratory Cooling Water Supply Cooling liquid and or chemicals Exhaust Line Gas Supplies Compressed Air Nitrogen Additional Gas Transpor...

Страница 22: ...l Zeiss Promenade 10 07745 Jena Germany phone 49 1803 33 63 34 fax 49 3641 64 3439 info microscopy de zeiss com www zeiss com microscopy Installation Requirements ZEISS LSM 900 en US Rev 2 Modificatio...

Отзывы: