2 Requirements | 2.4 Network Connectivity
2.4.1 ZEISS Predictive Service
ZEISS Predictive Service is used for condition monitoring by systematic retrieval of relevant instru-
ment health information and its server based processing. The goal is to detect deviations in the in-
strument performance before they impair the user’s result or even result in unplanned downtime.
For details, ask the ZEISS service representative about recommendations for data security policy
on ZEISS Systems.
For predictive service, we require the above network connectivity and predictive service will oper-
ate under the parameters shown in the table below.
Ports
443
Protocols
HTTPS, secure websocket
URL of ZEISS server
predictive-service.zeiss.com
IP addresses
52.174.243.245
Connection security
TLS 1.2 with AES 256 bit
Supported proxy settings for installation
§
No authentication
§
Basic authentication
§
Digest authentication
§
NTLM authentication
Available and patched local area network
connected to ZEISS system PC
RJ 45 connector for LAN
Security patched and measures according to
your local IT security standards
e. g. antivirus software, installed and main-
tained by customer/operator
Installation Requirements ZEISS LSM 900 | en-US | Rev. 2 | 349500-8183-000