2 Requirements | 2.4 Network Connectivity

ZEISS

2.3.2 UPS Requirements

If the site requires a UPS (Uninterruptable Power Supply), please ensure the KVA rating of the UPS
meets the ratings of the Microscope System and all third-party items that may also be connected,
e.g. water chiller, air compressor, sample prep equipment, EDS or WDS, etc.

This will ensure that the complete system will continue to function during a blackout or brownout
long enough to allow the user to save any experiments and data and to shut off the system prop-
erly.

As battery performance deteriorates over time, always overestimate the time (typically 2x) you re-
quire for the UPS to maintain power in a blackout (long mains interruption) or brownout (short
fluctuation of the mains supply) situation. The UPS only serves to bridge the time of a blackout or
brownout until normal mains is restored. Please ensure to switch off the Microscope System be-
fore the batteries run low in a blackout.

In areas where the mains supply is very unstable or not very clean, the most stable mode of UPS is
the online mode, where the UPS constantly generates the output mains supply.

Battery

Battery Charger

Inverter

LOAD

MAINS IN

Fig. 4: Block diagram of online UPS

Battery Charger

MAINS IN

Battery

Inverter

LOAD

Normal Mode

Backup Mode

Fig. 5: Block diagram of offline UPS

2.4 Network Connectivity

In order to provide remote diagnostics, ZEISS recommends connecting the computer to the inter-
net in a safe and secured method.

As a guideline, ask the ZEISS service representative about recommendations for data security pol-
icy on ZEISS Systems.

Cable type

CAT-6a

Connector type

RJ45 (IEC 60603-7 8P8C Modular connector)

Cable length

Site dependent – customer supplied. Not to
exceed total length of 100 m.

Protocol

Ethernet 10BASE-T/100BASE-TX/1000BASE-T

12

Installation Requirements ZEISS LSM 900 | en-US | Rev. 2 | 349500-8183-000

Содержание LSM 900

Страница 1: ...Installation Requirements ZEISS LSM 900 Laser Scanning Microscope...

Страница 2: ...archiving purposes shall remain unaffected thereby Any viola tions may be prosecuted as copyright infringements The use of general descriptive names registered names trademarks etc in this document d...

Страница 3: ...rvoltage Category 10 2 3 2 UPS Requirements 12 2 4 Network Connectivity 12 2 4 1 ZEISS Predictive Service 13 2 5 Safety and Radiation Information 14 2 5 1 Laser Safety 14 2 5 2 Biosafety Level of the...

Страница 4: ...oom temperature Air quality The customer is responsible for ensuring that the conditions specified in this document are met and that the required equipment is available before delivery and installatio...

Страница 5: ...than 50 dB Z class 200 to 500 Hz less than 55 dB Z class 2 1 2 Air Conditioning and Quality The following nominal power usage for the Microscope System and possible accessories is shown in Watts To m...

Страница 6: ...porary conductivity caused by condensation is to be expected Example A circuit used in an of fice environment The circuitry inside a computer would fall in this cate gory 3 Conductive pollution occurs...

Страница 7: ...d in closed rooms It is recommended to install the Microscope System in a dimmed room where artificial illumination sunlight or other light sources cannot interfere with image acquisition The Microsco...

Страница 8: ...depending on the options purchased with the system For possible changes in preferential layout please dis cuss with your ZEISS representative The placing of the user table can vary depending on avail...

Страница 9: ...ds and supplied socket with protective ground contact securely connected The provided electrical connection must be in accordance with the applicable electrical codes for the country of installation I...

Страница 10: ...eferred to as Installation Category to address transient voltages Category IV devices can handle the largest transients relative to the normal working voltage Category I devices can handle only small...

Страница 11: ...oltage levels of these loads must be rated at this overvoltage category III In fixed installations and for cases where the reliability and the availability of the equipment is subject to special requi...

Страница 12: ...til normal mains is restored Please ensure to switch off the Microscope System be fore the batteries run low in a blackout In areas where the mains supply is very unstable or not very clean the most s...

Страница 13: ...re the above network connectivity and predictive service will oper ate under the parameters shown in the table below Ports 443 Protocols HTTPS secure websocket URL of ZEISS server predictive service z...

Страница 14: ...s the responsible body the operator is responsible for complying with laser safety regula tions The following types of lasers are currently intended for use Using other lasers than indicated be low re...

Страница 15: ...ever some do exist in a small number of places in the US and around the world As the highest level of biological safety a BSL 4 lab consists of work with highly dangerous and exotic mi crobes Infectio...

Страница 16: ...e user if the air supplied is from the building facilities or from gas bottles Due to acoustic noise and vibrations where possible the compressor should be installed in a sep arate room close to the w...

Страница 17: ...f the system where necessary Some components such as the tables are large heavy or bulky and may require extra assis tance to get the units into the allocated site Maximum shock resistance Do not drop...

Страница 18: ...s used not possi ble for Airyscan Anti vibration plate for Axio Imager 320 300 45 7 Anti vibration plate for Axio Observer 800 525 45 7 Microscope Axio Examiner Z1 700 390 820 24 Axio Imager Z2 560 39...

Страница 19: ...EU RoHS and 2012 19 EU WEEE as applicable for the product In addition to the European and international guidelines and standards the 21 CFR 1040 10 Performance Standards for light emitting products la...

Страница 20: ...ch requirements are yet to be completed ZEISS can be requested to assist with the checking of the environmental conditions Environmental conditions in and around a system can change with time due to v...

Страница 21: ...adiation Information X Ray Laser Safety Biosafety Level of the Laboratory Cooling Water Supply Cooling liquid and or chemicals Exhaust Line Gas Supplies Compressed Air Nitrogen Additional Gas Transpor...

Страница 22: ...l Zeiss Promenade 10 07745 Jena Germany phone 49 1803 33 63 34 fax 49 3641 64 3439 info microscopy de zeiss com www zeiss com microscopy Installation Requirements ZEISS LSM 900 en US Rev 2 Modificatio...

Отзывы: