Operation
KSK 150 FC
Version 7.1
Page 54
G-30-1158-en
CAUTION
Risk of burns due to light intensity that is too high
R036
R095
Due to the use of an additional beam splitter in the main observer beam path,
less light is available to the user in the eyepieces. However, to obtain the
same brightness level in the image as provided without the additional beam
splitter, the user can increase the light intensity of the lighting equipment.
The patient will be exposed to increased illumination intensity. This can result
in a first degree burn.
• To maintain the loss of light for the main viewer at a minimum, to the
extent possible use only a beam splitter with a splitting ratio of 80:20 for
the camera.
• Set the illumination intensity as low as possible.
• Accordingly, allow greater illumination intensity to affect the patient only
briefly.
CAUTION
The illumination can be switched off and on automatically
The system monitors the temperature of the lamp. If the temperature is too
high, the light source switches off automatically. After the lamp has cooled
down, the device switches the light source back on automatically.
If a device is left unattended, long periods of illumination can result in burn
injuries.
• Always switch the device off when you are no longer using it.
R098
NOTE
Decrease in the lamp brightness or lamp failure!
• Check the lamp brightness before each use of the device.
• If the brightness fluctuates, exchange the lamp.
• Take appropriate precautions to ensure that the surgical procedure or
treatment can still be completed even if the full functionality of this col-
poscope is not available.
• If necessary, use an external OR illuminator to proceed.
R038
R039
NOTE
Interaction of heat and antimicrobial substances
High light intensity in the surgical field combined with a long treatment time
and the interaction of heat and antimicrobial substances in incision foils may
lead to an increase in the patient's reaction to these substances (skin irrita-
tion).
• Watch for such possible interaction during your treatments.
R053
Содержание KSK 150 FC
Страница 1: ...ZEISS Colposcope KSK 150 FC Instructions for Use G 30 1158 en Version 7 1 2015 06 11 ...
Страница 4: ...KSK150FC Version 7 1 Page 4 G 30 1158 en ...
Страница 15: ...Version 7 1 G 30 1158 en Page 15 KSK 150 FC Safety Measures ...
Страница 17: ...Version 7 1 G 30 1158 en Page 17 KSK 150 FC Safety Measures 1 5 1 1 1 6 2 2 4 2 3 ...
Страница 19: ...Version 7 1 G 30 1158 en Page 19 KSK 150 FC Safety Measures 7 10 8 11 12 9 8 ...
Страница 21: ...Version 7 1 G 30 1158 en Page 21 KSK 150 FC Safety Measures 19 18 17 13 14 15 16 ...
Страница 27: ...Version 7 1 G 30 1158 en Page 27 KSK 150 FC System Overview 1 3 2 1 3 4 2 5 6 7 ...
Страница 29: ...Version 7 1 G 30 1158 en Page 29 KSK 150 FC System Overview 3 4 5 6 7 1 2 8 9 10 ...
Страница 31: ...Version 7 1 G 30 1158 en Page 31 KSK 150 FC System Overview 1 3 6 2 5 4 8 10 11 9 12 7 ...
Страница 37: ...Version 7 1 G 30 1158 en Page 37 KSK 150 FC Preparations for use ...
Страница 45: ...Version 7 1 G 30 1158 en Page 45 KSK 150 FC Preparations for use 1 2 ...
Страница 63: ...Version 7 1 G 30 1158 en Page 63 KSK 150 FC Care and Maintenance 4 1 5 3 2 ...
Страница 66: ...Care and Maintenance KSK 150 FC Version 7 1 Page 66 G 30 1158 en ...
Страница 84: ...System Data KSK 150 FC Version 7 1 Page 84 G 30 1158 en ...
Страница 90: ...Index KSK 150 FC Version 7 1 Page 90 G 30 1158 en ...
Страница 91: ...Version 7 1 G 30 1158 en Page 91 KSK 150 FC Blank page for your notes ...