Preparations for use
KSK 150 FC
Version 7.1
Page 46
G-30-1158-en
Connecting the system
Connecting the device cable
Make sure that the device has been switched off at the power switch before
plugging in or unplugging the device cable.
• Connect the device cable to the device by inserting connector (2) into
socket (4) on the device.
• Connect the device to line power by inserting the power plug (3) into a
socket of the examination room.
Only use a power outlet which is provided with a properly connected pro-
tective earth conductor.
CAUTION
Increased leakage current!
R046
There is a risk of user and patient sustaining an electric shock.
• Never touch the video plugs (connectors) while in contact with the pa-
tient.
• Do not connect any defective or unapproved accessories to the contacts
of connectors.
CAUTION
Risk of tripping over cables lying on the floor!
R057
If cables are inappropriately installed on the floor, this increases the risk of
tripping and interferes with the flow of work. By tripping and falling down,
users may suffer tissue trauma.
• If possible, cables should only be laid outside of the area of movement.
NOTE
Observe the rated voltage
Prior to leaving the factory, the device is adjusted to the rated voltage of the
respective country of destination. The set voltage is visible in a small
window (1) on the carrier arm. If the set voltage does not comply with the
line voltage available at the site of installation, this can cause damage to the
device.
• Check the voltage setting. The set voltage must comply with the line volt-
age available at the site of installation. If this is not the case, you must
not operate the system.
• If there is a variance, ZEISS Service will have to convert your device to the
line voltage available at your location.
R017
Содержание KSK 150 FC
Страница 1: ...ZEISS Colposcope KSK 150 FC Instructions for Use G 30 1158 en Version 7 1 2015 06 11 ...
Страница 4: ...KSK150FC Version 7 1 Page 4 G 30 1158 en ...
Страница 15: ...Version 7 1 G 30 1158 en Page 15 KSK 150 FC Safety Measures ...
Страница 17: ...Version 7 1 G 30 1158 en Page 17 KSK 150 FC Safety Measures 1 5 1 1 1 6 2 2 4 2 3 ...
Страница 19: ...Version 7 1 G 30 1158 en Page 19 KSK 150 FC Safety Measures 7 10 8 11 12 9 8 ...
Страница 21: ...Version 7 1 G 30 1158 en Page 21 KSK 150 FC Safety Measures 19 18 17 13 14 15 16 ...
Страница 27: ...Version 7 1 G 30 1158 en Page 27 KSK 150 FC System Overview 1 3 2 1 3 4 2 5 6 7 ...
Страница 29: ...Version 7 1 G 30 1158 en Page 29 KSK 150 FC System Overview 3 4 5 6 7 1 2 8 9 10 ...
Страница 31: ...Version 7 1 G 30 1158 en Page 31 KSK 150 FC System Overview 1 3 6 2 5 4 8 10 11 9 12 7 ...
Страница 37: ...Version 7 1 G 30 1158 en Page 37 KSK 150 FC Preparations for use ...
Страница 45: ...Version 7 1 G 30 1158 en Page 45 KSK 150 FC Preparations for use 1 2 ...
Страница 63: ...Version 7 1 G 30 1158 en Page 63 KSK 150 FC Care and Maintenance 4 1 5 3 2 ...
Страница 66: ...Care and Maintenance KSK 150 FC Version 7 1 Page 66 G 30 1158 en ...
Страница 84: ...System Data KSK 150 FC Version 7 1 Page 84 G 30 1158 en ...
Страница 90: ...Index KSK 150 FC Version 7 1 Page 90 G 30 1158 en ...
Страница 91: ...Version 7 1 G 30 1158 en Page 91 KSK 150 FC Blank page for your notes ...