Operation
Humphrey FDT User Manual
2660021149532 Rev. B 2015-11
3-14
Because the FDT Screening N-30-1 (or C-20-1) and N-30-5 (or C-20-5) programs utilize
different baseline normal contrast levels, they also differ in how they flag (shade) test
locations as being outside normal limits. With the N-30-1 (or C-20-1) test, points are
initially flagged when they reach the P<1% probability significance level (i.e. the sensitivity
level which is found less than 1% of the time in the normal population). With the N-30-5
(or C-20-5) test, points are initially flagged when they reach the P<5% probability
significance level (i.e. the sensitivity level which is found less than 5% of the time in the
normal population), and thus there is a greater chance that any point may be deemed
outside normal limits. Therefore, the N-30-5 (or C-20-5) test is slightly more sensitive,
readily identifying more extensive loss. Below is a clinical example that demonstrates this
difference with a comparison between the Screening N-30-1 and N-30-5 programs.
Содержание Humphrey FDT 710
Страница 1: ...Humphrey FDT Model 710 User Manual...
Страница 19: ...Humphrey FDT User Manual 2660021149532 Rev B 2015 11 Introduction 2 7 Patient Video Screen Patterns...
Страница 20: ...Introduction Humphrey FDT User Manual 2660021149532 Rev B 2015 11 2 8...
Страница 38: ...Operation Humphrey FDT User Manual 2660021149532 Rev B 2015 11 3 18 Screening C 20 Test Result Sample...
Страница 39: ...Humphrey FDT User Manual 2660021149532 Rev B 2015 11 Operation 3 19 Threshold C 20 Test Results Sample...
Страница 40: ...Operation Humphrey FDT User Manual 2660021149532 Rev B 2015 11 3 20 Threshold N 30 Test Results Sample...
Страница 50: ...Service Information Humphrey FDT User Manual 2660021149532 Rev B 2015 11 5 2...
Страница 53: ......