Humphrey FDT User Manual
2660021149532 Rev. B 2015-11
Safety Information
1-1
(1) Safety Information
All operating personnel should be familiar with the general safety information in this
summary. Additional safety information may also be found throughout this manual.
Note: If a serious incident has occurred in relation to this medical device, to the user, or to
another person, then the user (or responsible person) must report the serious incident to
the medical device manufacturer or the distributor. In the European Union, the user (or
responsible person) must also report the serious incident to the Competent Authority in the
state where the user is established.
Summary
CAUTION: Before connecting the power cord to the appliance inlet, verify that the
voltage selector switch is correctly set (115 V or 230 V) to match the power
requirements in your region.
CAUTION: Always replace fuses with the same type and rating (T 0.400 A 250 V).
Refer to Fuse Replacement section for instruction on changing fuses.
CAUTION: Do Not Sterilize the instrument or any of its components.
CAUTION: After unpacking the instrument, pull down the printer door (below the
LCD display), using the finger cutouts on the sides of the door (near the top), and
remove the foam shipping wedge before using the printer. Close the printer door. Be
sure the paper is sticking out through the slot in the door. Failure to remove the
shipping wedge will result in improper operation of the printer.
CAUTION: Do not use the instrument near other equipment which produces strong
magnetic fields (such as MRI). The video monitor performance may be adversely
affected.
Product Safety
This instrument is classified as follows:
• Class I Equipment – Protection against electrical shock.
• Type BF – Degree of protection against electric shock of applied part (forehead rest
and Patient Response button).
• Ordinary Equipment (IPX0) – Degree of protection against ingress of liquids (none).
• Continuous Operation – Mode of operation.
WARNING: This device contains visual stimuli, including flickering light
and flashing patterns, between 5 and 65 Hz. Medical professionals need to
determine whether this device should be used for patients who may be
photosensitive, including those with epilepsy.
WARNING: To prevent electric shock, the instrument must be plugged into
an earthed ground outlet. Do not remove or disable the ground pin.
Содержание Humphrey FDT 710
Страница 1: ...Humphrey FDT Model 710 User Manual...
Страница 19: ...Humphrey FDT User Manual 2660021149532 Rev B 2015 11 Introduction 2 7 Patient Video Screen Patterns...
Страница 20: ...Introduction Humphrey FDT User Manual 2660021149532 Rev B 2015 11 2 8...
Страница 38: ...Operation Humphrey FDT User Manual 2660021149532 Rev B 2015 11 3 18 Screening C 20 Test Result Sample...
Страница 39: ...Humphrey FDT User Manual 2660021149532 Rev B 2015 11 Operation 3 19 Threshold C 20 Test Results Sample...
Страница 40: ...Operation Humphrey FDT User Manual 2660021149532 Rev B 2015 11 3 20 Threshold N 30 Test Results Sample...
Страница 50: ...Service Information Humphrey FDT User Manual 2660021149532 Rev B 2015 11 5 2...
Страница 53: ......