Safety Information
Humphrey FDT User Manual
2660021149532 Rev. B 2015-11
1-6
Guidance and Manufacturer’s Declarations
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The Humphrey FDT is intended for use in the electromagnetic environment specified below. The customer or user of the
Humphrey FDT should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF Emissions
CISPR 11
Group 1
The Humphrey FDT uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
RF Emissions
CISPR 11
Class B
The Humphrey FDT are suitable for use in all establishments including
domestic and those directly connected to the public low-voltage power sup-
ply network that supplies buildings used for domestic purposes.
Harmonics
IEC 61000-3-2
Class A
Flicker
IEC 61000-3-3
Complies
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Humphrey FDT is intended for use in the electromagnetic environment specified below. The customer or user of the
Humphrey FDT should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic Environment -
Guidance
ESD
IEC 61000-4-2
± 6kV Contact
± 8kV Air
± 6kV Contact
± 8kV Air
Floors should be wood, concrete, or ceramic
tile. If floors are synthetic material, the relative
humidity should be at least 30%.
EFT
IEC 61000-4-4
±2kV Mains
±1kV I/O
±2kV Mains
Not Applicable
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±1kV Differential
±2kV Common
±1kV Differential
±2kV Common
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage Dips, short
interruptions, and
voltage variations on
power supply input
lines.
IEC 61000-4-11
>95% Dip in 0.5
Cycle
60% Dip in 5 Cycles
30% Dip for 25
Cycles
>95% Dip in
5 Seconds
>95% Dip in 0.5
Cycle
60% Dip in 5 Cycles
30% Dip for 25
Cycles
>95% Dip in
5 Seconds
Mains power quality should be that of a typical
commercial or hospital environment. If the
user of the Humphrey FDT requires continued
operation during power mains interruptions, it
is recommended that the Humphrey FDT be
powered from an uninterruptible power supply
or battery.
Power Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
Содержание Humphrey FDT 710
Страница 1: ...Humphrey FDT Model 710 User Manual...
Страница 19: ...Humphrey FDT User Manual 2660021149532 Rev B 2015 11 Introduction 2 7 Patient Video Screen Patterns...
Страница 20: ...Introduction Humphrey FDT User Manual 2660021149532 Rev B 2015 11 2 8...
Страница 38: ...Operation Humphrey FDT User Manual 2660021149532 Rev B 2015 11 3 18 Screening C 20 Test Result Sample...
Страница 39: ...Humphrey FDT User Manual 2660021149532 Rev B 2015 11 Operation 3 19 Threshold C 20 Test Results Sample...
Страница 40: ...Operation Humphrey FDT User Manual 2660021149532 Rev B 2015 11 3 20 Threshold N 30 Test Results Sample...
Страница 50: ...Service Information Humphrey FDT User Manual 2660021149532 Rev B 2015 11 5 2...
Страница 53: ......