Humphrey FDT User Manual
2660021149532 Rev. B 2015-11
Operation
3-11
False Negative and False Positive catch trial presentations are indicated by the symbols (+)
& (-) on the operator LCD so they can be distinguished from regular teststimulus
presentations.
Note: You can
PAUSE
or
RE-START
the test by pressing the
GREEN Button
at any time
during the test. Follow the Operator LCD Display instructions to
CONTINUE TEST
or to
RE-START TEST
from a
pause
. The patient can also PAUSE the test (for a break, etc.) by
simply pressing and holding down the
Patient Response Button
. The test will resume
automatically once the patient releases the
Patient Response Button.
During patient
pause, you may also select
OPERATOR PAUSE (GREEN Button)
so that the test will
remain paused until you restart it.
Note
:
A
SCREENING C-20 TEST
takes less than
1 minute
(per eye) to complete, a
THRESHOLD C-20 TEST
takes approximately
4 minutes
(per eye) to complete and a
THRESHOLD N-30 TEST
takes about 4-1/2 minutes.
At the end of the right eye test, the Operator LCD Display will prompt for a left eye test.
Slide the Patient Visor to the left eye test position (this is to your left when facing the
patient from the operator’s side). Prepare the patient. Select RUN SCREENING C-20 TEST
(TOP Operator Button), RUN THRESHOLD C-20 TEST (2nd Operator Button from the top), or
RUN THRESHOLD N-30 TEST (3rd Operator Button from the top) from the LEFT EYE TEST
MENU to begin the left eye test immediately after a momentary check for proper
calibration.
Note: To skip the left
eye test, select
SKIP LEFT EYE
(Bottom Operator Button) to
proceed to the results menu.
LEFT EYE TEST MENU
RUN FULL THRESHOLD C-20 TEST
SKIP LEFT EYE
GRN BUTTON CANCELS
24 JUN 1997 04:10 PM
Содержание Humphrey FDT 710
Страница 1: ...Humphrey FDT Model 710 User Manual...
Страница 19: ...Humphrey FDT User Manual 2660021149532 Rev B 2015 11 Introduction 2 7 Patient Video Screen Patterns...
Страница 20: ...Introduction Humphrey FDT User Manual 2660021149532 Rev B 2015 11 2 8...
Страница 38: ...Operation Humphrey FDT User Manual 2660021149532 Rev B 2015 11 3 18 Screening C 20 Test Result Sample...
Страница 39: ...Humphrey FDT User Manual 2660021149532 Rev B 2015 11 Operation 3 19 Threshold C 20 Test Results Sample...
Страница 40: ...Operation Humphrey FDT User Manual 2660021149532 Rev B 2015 11 3 20 Threshold N 30 Test Results Sample...
Страница 50: ...Service Information Humphrey FDT User Manual 2660021149532 Rev B 2015 11 5 2...
Страница 53: ......