Maintenance
Instructions for Use
Every 12 months
Component
Procedure / Criteria
Documentation and
identification labels
Visual inspection:
Manuals and identification labels available, undamaged, complete and
legible
Carrier system
including microscope
mount
Function test:
−
Play-free movement of bearings, stops and suspensions
−
Locking screws are in place and tightened
−
Movement of the gas pressure spring is homogeneous and jolt-free
and
the gas pressure spring is not operated in the limit range
−
Brakes operational at maximum permissible OPMI configuration
Stand column (floor
stand only)
Function test:
Secure seating without mechanical play in the column connection
Casters (floor stand
only)
Visual inspection and function test:
Play-free seating, no locking during rotating and rolling, locking tabs and
brakes securely positioned
OPMI
Function test:
Check of optical imaging and illumination of the field of view
Focus
Function test:
Play-free adjustability of focus
Varioskop, tube and
other accessories
mounted on the OPMI
(camera, etc.)
Function test:
Attached parts without mechanical play, knurled screws tightened and
other removable microscope components securely mounted
Every 2 years
Component
Procedure / Criteria
Filters of the filter
wheel assembly inside
OPMI
Check the filter wheel assembly
Every 4 years
Component
Procedure / Criteria
Carrier system
including microscope
mount
Visual inspection:
−
No signs of wear on the supporting structure
−
No signs of wear or damage on the interfaces and couplings
Cables, plugs and
switches
Visual inspection:
−
Undamaged insulation and connectors on system/cable
−
Check cable guide including fasteners
Casters of floor stand
Replace the casters if the outer rubber ring starts to detach
39 / 59