Instructions for Use
Maintenance
Safety check
WARNING!
Risk of injury!
Cables and plug connections may come loose and parts of the housing
may become live. The user and patient may suffer an electric shock.
•
To prevent any impairment of the system's safety as a result of
aging, wear etc., the operating company must ensure, in
accordance with the applicable national regulations, that the
regular technical safety checks defined for this system are
performed on schedule and to the stipulated extent.
The safety checks may only be carried out by the manufacturer or by
qualified persons.
The scope of the safety checks of the system should comprise at least the
following points:
‒
Availability of the Instructions for Use
‒
Visual inspection of system and accessories for damage and legibility
of labels
‒
Leakage current test
‒
Test of protective ground conductor
‒
Function and wear test of casters and brakes (locking knobs)
‒
Function test of all switches, buttons, sockets and indicator LED of
the system
‒
Securing screws are in place and tightened
!
305935-3500-000 – 2.1 – 2017-12-15
40 / 59
Содержание EXTARO 300
Страница 1: ...ZEISS EXTARO 300 Instructions for Use...