Notes on device safety
Colibri 7
6
423052-7344-001
02/2017
Depending on the design of the microscope, the risk for the user must be assessed as being markedly
lower.
When it comes to the intended use of the Axio Observer microscope with Colibri 7, for example, the risk
status of the overall system is downgraded to LED risk group 2 according to DIN EN 62471:2009.
Colibri 7 and its original accessories may only be used for the purposes described in this operating
manual.
Particular attention should be paid to the following:
Use Colibri 7 for ZEISS microscopes only. The manufacturer cannot assume any liability for any
other applications of the instrument, including the use of individual modules or components.
This also applies to any service or repair work that is not carried out by authorized service
personnel. Failure to comply with this shall render all warranty claims invalid.
To ensure safe operation of the Colibri 7 solid-state light source in connection with the
microscope, pay attention to the operating manual of the microscope.
Colibri 7 is not to be used in explosive environments.
Keep combustible and easily inflammable materials away from the light beam and its
surroundings.
Colibri 7 is not equipped with any special devices for protection from substances that are
corrosive, potentially infectious, toxic, and radioactive or other substances that may be
hazardous to health. Observe all legal regulations, particularly the relevant national accident
prevention regulations when handling such substances.
Disconnect the power plug of the desktop power supply unit prior to assembly / disassembly of
the lamp module.
Only specially authorized experts or service staff are permitted to open the lamp module.