OPERATION
4.1
Starting operation of Colibri 7 for the first time
Colibri 7 does not have a separate on / off switch.
After mounting it to a microscope it only has to be connected to a power outlet.
•
Connect the Colibri 7 desktop power supply unit (Fig. 10/
7
) with a power supply cable to a power
outlet.
After a short initialization time Colibri 7 is ready for operation.
•
Switch on the microscope.
•
Start the control software used (from ZEN 2.3 (blue edition) or third-party software).
If the control panel or the LED Trigger Box are connected, they will be powered via Colibri 7 and are also
ready for use.
4.2
Controlling the Colibri 7 solid-state light source
Colibri 7 can be controlled in various ways depending on the available components:
−
via TFT display and selector wheel for illumination intensity at the Axio Observer, Axio Imager or
Axio Examiner microscopes, see section 4.2.1
−
via the control panel for Colibri 7, see section 4.2.2
−
via PC and installed software ZEN 2.3 (blue edition), see section 4.2.3
−
via LED Trigger Box or external triggering via third-party software