6-8
Start-up
61211-1020202
CenterMax Operating Instructions
Checking the protective circuit
The CMM is equipped with a protective circuit which monitors drive-
forces during bridge positioning movements in the x and y axes. The
shearing force monitoring is effective in all operating modes.
Function of the protective circuit
Drives must cut out
automatically.
If the traveling motion is obstructed by a collision in the x or y axis and
the shearing force therefore exceeds a value of approx. 150N, the
drives must cut off automatically.
Danger!
In order to ensure the safety of the CMM during measurement, the
function of the protective circuit must be checked regularly.
•
Check the protective circuit on a
monthly
basis.
Check
When performing the check, you must proceed as follows:
x axis
• Move the crossbeam forward just far enough so that you can
grasp the quill housing with one hand.
!
Содержание CenterMax
Страница 1: ...Operating Instructions CenterMax Coordinate measuring machine for series production...
Страница 9: ...61211 1020202 Operating Instructions Contents7 Appendix Application examples 1...
Страница 10: ...61211 1020202 Operating Instructions Contents 8...
Страница 14: ...Foreword 61211 1020202 Operating Instructions...
Страница 26: ...1 12 Introduction 61211 1020202 CenterMax Operating Instructions...
Страница 58: ...3 6 Technical data 61211 1020202 CenterMax Operating Instructions...
Страница 64: ...4 6 Transport and installation 61211 1020202 CenterMax Operating Instructions...
Страница 80: ...5 16 Preparations for start up 61211 1020202 CenterMax Operating Instructions...
Страница 100: ...6 20 Start up 61211 1020202 CenterMax Operating Instructions...
Страница 134: ...7 34 Measuring operation 61211 1020202 CenterMax Operating Instructions...
Страница 144: ...8 10 Errors and faults 61211 1020202 CenterMax Operating Instructions...
Страница 156: ...9 12 Maintenance and care 61211 1020202 CenterMax Operating Instructions...
Страница 162: ...61211 1020202 Operating Instructions Glossary 2...
Страница 164: ...61211 1020202 Operating Instructions Appendix 2...
Страница 165: ......
Страница 166: ...01 02 61211 1020202 Operating Instructions...