7-4
Measuring operation
61211-1020202
CenterMax Operating Instructions
Probing with VAST
The probe must be moved back counter to the probing direction after
each single point probing.
Measurement in static
status
The measuring data are not accepted immediately after probing.
Instead, they are initially polled following a damping period of approx.
one second. I.e. the actual measurement occurs only after the
machine reaches a static state.
Multipoint measurement
Multipoint measurement is only possible with the VAST probe system.
During multipoint measurement, the probe is moved perpendicular to
the probing direction.
The probing axis runs parallel to the workpiece surface. During mea-
surement, the probe element constantly remains in contact with the
workpiece surface. The measuring force remains effective until prob-
ing is completed.
A
Lowering the probe and probing the workpiece; with the left joystick.
B
-
D
Move the probe over the workpiece with the right joystick. The
position in the z direction remains unchanged.
D
Lifting the probe.
A
B
D
A
C
B
Содержание CenterMax
Страница 1: ...Operating Instructions CenterMax Coordinate measuring machine for series production...
Страница 9: ...61211 1020202 Operating Instructions Contents7 Appendix Application examples 1...
Страница 10: ...61211 1020202 Operating Instructions Contents 8...
Страница 14: ...Foreword 61211 1020202 Operating Instructions...
Страница 26: ...1 12 Introduction 61211 1020202 CenterMax Operating Instructions...
Страница 58: ...3 6 Technical data 61211 1020202 CenterMax Operating Instructions...
Страница 64: ...4 6 Transport and installation 61211 1020202 CenterMax Operating Instructions...
Страница 80: ...5 16 Preparations for start up 61211 1020202 CenterMax Operating Instructions...
Страница 100: ...6 20 Start up 61211 1020202 CenterMax Operating Instructions...
Страница 134: ...7 34 Measuring operation 61211 1020202 CenterMax Operating Instructions...
Страница 144: ...8 10 Errors and faults 61211 1020202 CenterMax Operating Instructions...
Страница 156: ...9 12 Maintenance and care 61211 1020202 CenterMax Operating Instructions...
Страница 162: ...61211 1020202 Operating Instructions Glossary 2...
Страница 164: ...61211 1020202 Operating Instructions Appendix 2...
Страница 165: ......
Страница 166: ...01 02 61211 1020202 Operating Instructions...