NOTES ON INSTRUMENT SAFETY
Cell Observer SD
Safety Instructions
Carl Zeiss
M60-2-0021 e 01/2009
423638-7044-001
9
On completion of work for installation or retrofitting of the Cell Observer SD System, an
authorized expert must perform a careful inspection to check for absence of defects. He/she
must also make sure that all laser protection shields are properly mounted.
Tube openings and other unused connection terminal must always be protected from dust and
moisture with appropriate sealing aids or anti-dust caps.
Do not operate the equipment delivered in potentially explosive atmospheres.
The Cell Observer SD is only approved for operation in closed rooms.
Establish a suitable workplace environment in order to prevent the formation of electrostatic
charges in electronic components.
Always disconnect the instrument from the power outlet before opening it and changing the
fuses.
Only use fuses appropriate for the rated current as specified. The use of makeshift fuses and
short-circuiting of the fuse holders are not permitted.
On completion of work for installation or conversion of the Cell Observer SD System, an
authorized expert must perform a careful inspection to check for proper working condition.
He/she must also make sure that all laser protection shields are correctly mounted.
Tube openings and other unused connection terminals must always be protected from dust and
moisture with appropriate sealing aids or anti-dust caps.
Dust and dirt may impair the instrument’s performance. The instrument must therefore be
protected as far as possible from such influences.
Clogged or covered ventilation slats may lead to heat build-up that will damage the instrument
and, in extreme cases, cause a fire. Always keep the ventilation slats clear and ensure that no
objects enter the instrument through the ventilation slats.
To prevent vibrations during Cell Observer SD operation, you should work with an Axio
Observer anti-vibration plate.