Intended Use
Cell Observer SD
20
423638-7044-001
M60-2-0021 e 01/2009
2
INSTRUMENT DESCRIPTION
2.1
Intended Use
Product name:
CSU-X1M 1800 spinning disc unit
CSU-X1M 1800 Dual Cam spinning disc unit
CSU-X1M 5000 spinning disc unit
CSU-X1M 5000 Dual Cam spinning disc unit
CSU-X1A 5000 spinning disc unit
When configured with a CSU-X1 spinning disc unit, the Cell Observer SD provides a system to capture
confocal fluorescence pictures in real time. This is achieved with a multi-hole Nipkow scanning disc. A
multi-laser module is used as the light source. Alternatively, a multi-line Ar-laser may be selected.
One or two CCD cameras perform real-time image capturing.
The spinning disc unit should only be operated when combined with an Axio Observer.Z1 microscope
stand.
Human blood and tissue samples may be examined.
The laser modules (refer to configuration options, section 2.4 on page 30) may not be operated
on the Cell Observer SD, unless combined with a basic safety kit for Z1 "Axio Observer"
microscope stand (432927-9010-000).
Use the Cell Observer SD only with laser class 3B laser modules up to 300 mW of output
power.
Carl Zeiss MicroImaging GmbH cannot warrant the proper function of microscope system
components or the compliance with laser system safety requirements if lasers of a higher power
rating are selected.