START-UP
Cell Observer SD
Basic Microscope Settings
Carl Zeiss
M60-2-0021 e 01/2009
423638-7044-001
41
3.3
Basic Microscope Settings
Installation and commissioning of the Cell Observer SD system may not be performed by
anyone other than service personnel of Carl Zeiss MicroImaging GmbH or technicians duly
authorized to handle such work by the Manufacturer.
•
Connect the microscope to your computer (if not yet connected).
•
Install AxioVision Rel. 4.7 image processing software on your computer (unless already installed).
•
Mount objective lenses into the lens turret as described in the
Axio Observer Operating Manual
and
keep Immersol 518 F
®
ready for this purpose.
Read in any case the Immersol 518 F
®
safety data sheet.
•
Turn the lens turret into an empty position.
3.4
Height Setting of CSU-X1 Spinning Disc Unit
Before the CSU-X1 spinning disc unit can be mounted onto the microscope, the standard
support legs of the Axio Observer must be replaced with special version legs. This will raise the
unit by 5 mm.
For replacement of support legs, you should refer to the Axio Observer Operating Manual.
•
Screw the C-mount camera adapter to the
spinning disc unit.
•
Use two fixing screws (Fig. 3-3/
1
) to attach the
support leg assembly to the camera adapter of
the spinning disc unit.
•
Turn setting screws (Fig. 3-3/
2
) (3" Allen key) as
necessary to adjust the vertical leg position. Use
the ring nuts (Fig. 3-3/
3
) to lock the support
legs in position.
Fig. 3-3
Mounting the support legs