OPERATION
Cell Observer SD
Switching the System on
Carl Zeiss
M60-2-0021 e 01/2009
423638-7044-001
53
4.2.2
Turning the Multiline Argon Laser On
•
Make sure that the toggle switch (Fig. 4-5/
3
) at the laser power pack control module (Fig. 4-5/
1
) is in
Standby
position (down) and the power controller (Fig. 4-5/
2
) in left stop position before you turn the
laser on.
•
Transfer the On/Off switch at the power pack into position
I
3
). Power is supplied to the
argon laser.
•
Turn the key switch at the power pack into position
I
2
) to trigger the argon laser.
1
Casing
2
Potentiometer for power control
3
Standby
switch
4
Green
LED
optimal
(lights when laser energy within
nominal
range)
5
Red
LED
reduced life time
(lights when laser energy
too
high)
Fig. 4-5
Control unit of multiline argon laser
1
Power supply unit of 100 mW laser
2
Key
switch
3
On/Off
switch
Fig. 4-6
Power supply unit of multiline argon
laser
Nominal laser power will be available within a few seconds. Operation at nominal power and
the specified laser lifetime can only be warranted if the power selection potentiometer
(Fig. 4-5/
2
) is in a position approximately 12 hours. The green LED (Fig. 4-5/
4
) will light in this
case. For operation at a lower power setting, the potentiometer can be rotated anticlockwise. If
operated with a power setting above nominal level (potentiometer in a position greater than 12
hours), the laser’s lifetime will be reduced. The red LED will light (Fig. 4-5/
5
).
If the safety device is found to function in a non-conforming manner, you should shut the
system down, clearly mark it as defective and contact the Service Department of Carl Zeiss
MicroImaging GmbH immediately.