START-UP
Axiolab 5
Mounting standard components
ZEISS
05/2019
430037-7444-001
69
•
Loosen the three side screws (Fig. 3-24/
4a
,
4b
,
4c
) with an Allen wrench (AF 2.5) (Fig. 3-24/
3
) and
remove the old LED illuminator (Fig. 3-24/
2
) from the adapter (Fig. 3-24/
1
).
•
Insert the new LED illuminator into the adapter and tighten the three side screws on the adapter.
•
Insert the new LED illuminator with adapter (Fig. 3-23/
4
) into the holding tube (Fig. 3-23/
1
) up to the
stop.
•
Position the new LED illuminator with the pinhole (Fig. 3-23/
3
) at the left, or with the adapter front
edge (Fig. 3-23/
2
) aligned to the right.
•
Plug the lamp plug (Fig. 3-23/
5
) into the connector of the stand (Fig. 3-23/
6
).
•
Lift the securing clips (Fig. 3-22/
1
) of the lamp holder until they are completely enclosed in the lamp
holder. In doing so, slightly press the ends of the securing clips together and guide past the two upper
retaining elements. Release the pressure until the clamping bracket opens and engages on both sides
in the retaining elements.
Fig. 3-24
Changing the LED illuminator in the adapter
Reinserting the cover
•
Insert the cover (Fig. 3-20/
2
) at a slant from below into the upper holding elements of the stand,
swivel down and press into place.
•
Tighten screw (Fig. 3-20/
1
).
•
Reconnect the power cord.