START-UP
Axiolab 5
Mounting standard components
ZEISS
05/2019
430037-7444-001
53
3.1.2
Attaching the binocular tube/photo tube
For tubes mounted
without
an intermediate plate, proceed as follows:
•
Loosen the screw (Fig. 3-3/
4
) with an Allen wrench (AF 3). Remove the dust caps (Fig. 3-3/
2
,
5
) from
the underside of the tube and the dovetail ring mount on the stand side.
•
Hold the binocular tube/photo tube (Fig. 3-3/
1
) at an angle, insert it with the dovetail ring into the
stand mount (Fig. 3-3/
3
) and turn into a horizontal position. Rotate the binocular tube into the desired
observation position and re-tighten the Allen screw with the Allen wrench.
Fig. 3-3
Attaching the binocular tube
For tubes mounted
with
an intermediate plate on the Axiolab 5 stand, Pol-TL/conoscopy (430037-
90042-000), proceed as follows:
•
Loosen the screw (Fig. 3-3/
6
) with an Allen wrench (AF 3). Remove dust caps (Fig. 3-3/
8
,
11
) from the
underside of the tube and the dovetail ring mount on the stand side.
•
Insert the intermediate plate (Fig. 3-3/
10
) with its dovetail ring into the stand mount (Fig. 3-3/
7
) and
tighten Allen screw (Fig. 3-3/
6
).
•
Insert the binocular tube/photo tube (Fig. 3-3/
12
) into the intermediate plate, align and tighten the
screw (Fig. 3-3/
9
) with the Allen wrench.