40
Other Guidelines
Using the STOP button
The STOP button is for instantaneous deactivation of a moving motorized focusing
drive to prevent collision with a specimen on the latter.
1.
Press the
STOP
button to switch off the focusing drive. The
STOP
button is on the
left-hand side of the drive
2.
The
STOP
button engages and the focusing movement is immediately
interrupted
3.
To unlock the
STOP
button, pull it out again. The
STOP
button must be unlocked
to restart the focusing drive
Operating Manually
1.
Place a specimen in the centre of the round insert plate
2.
Pay attention to the following operation modes:
a.
To use only the incident light and respective filter sets, operate solely with
the Fluar illuminator Z mot and make sure the diaphragm slide is closed
b.
Alternatively, to use only the transmitted light, press the number 4 in the
Fluar illuminator Z mot and operate with the diaphragm slide and setting wheel
c.
Finally, to switch between lights, operate with both Fluar illuminator Z
mot, setting wheel (See Other Guidelines -> Contrast Methods and Filter Sets) and
diaphragm slide