OPERATION
Carl Zeiss
Coded Axio Imager operation and function controls
Axio Imager
80
430000-7344-001
M70-2-0020 e 06/2009
Intensity button (4-1/34)
−
Reduces the light intensity of halogen illuminator HAL.
Intensity button (4-1/35)
−
Increases the light intensity of halogen illuminator HAL.
Toggle switch for transmitted/reflected-light halogen illuminator (4-1/36)
−
Alternately switches the halogen illuminator for transmitted light or reflected light on/off.
−
Toggle switch up:
Reflected-light halogen illuminator ON (transmitted-light OFF)
−
Toggle switch down:
Transmitted-light halogen illuminator ON (reflected-light OFF)
−
Saves the values of the current light intensity and the optical path nosepiece position (brief beep).
−
Adaptation of the luminosity of the ring-like arranged LEDs at the light intensity control:
−
Switch on the illuminator and adjust light intensity at the control so that all LEDs light up.
−
Maintain
LM-Set
key pressed down. You hear a long beep tone.
−
Adjust the desired light intensity using the light intensity control (4-1/
14
).
−
Release the
LM-Set
key to save the light intensity adjusted.
Lever for reflected-light diffusing glass (4-1/38)
−
Lever in top position:
Diffusing glass in light path
−
Lever in bottom position: Diffusing glass removed from light path