![Zeiss Axio Imager Скачать руководство пользователя страница 73](http://html1.mh-extra.com/html/zeiss/axio-imager/axio-imager_operating-manual_3338713073.webp)
START-UP
Axio Imager
Installing Polarizer D, fixed
Carl Zeiss
M70-2-0020 e 06/2009
430000-7344-001
73
3.35
Installing Polarizer D, fixed
If the stand is provided with a
detachable stage carrier, remove it (see
Section 3.13), put it aside upside down
and install the polarizer from top.
•
Remove any installed polarizer or color filter
carrier from the condenser carrier.
•
Fasten adapter plate (3-59/
1
) with screws
2
) to the condenser carrier making use of
the provided holes (3-59/
6
and
7
).
•
Hold the polarizer (3-59/
4
) parallel to the
bottom side of the condenser carrier. Using the
offset adjusting lever (3-59/
3
), screw the
retaining bolt of the polarizer (3-59/
4
) into the
rear tapped hole of the adapter plate as far as it
will go.
•
Using the adjusting lever (3-59/
3
), screw the
detent bolt (3-59/
5
) into the front tapped hole
of the adapter plate as far as it will go.
For the installation of other polarizers or of the
color glass holder, proceed analogously.
Fig. 3-59
Installing Polarizer D, fixed