Page 28
W25 Analog Output
Offset
Allows having an offset of gloss units on the analog
output. The default value is 0. When you give an Offset
of e.g. 10, the analog output will stay at 0V (in 0-10V
mode) or 4 mA (in 4-20mA mode) until 10 gloss units
have been reached.
W26 Analog Output
Ceiling
Sets the gloss value when the analog output is at
100%. The default value is 100. This means when 100
gloss units are measured, the analog output will give
out 10V (in 0-10V mode) or 20 mA (in 4-20mA mode).
W27 Analog Output
Type
Three analog output modes are available:
0-10V voltage mode
0-20 mA current loop mode
4-20 mA current loop mode
W28 Hide gloss value
when not
precisely on
measuring
position
When this function is enabled, the gloss value on the
graph will only be updated while the measuring head
is precisely on the measuring position. In all other
cases, the measured value will not be updated.
W29 Always move to
measuring
position after
calibration
When this function is enabled, the measuring head
moves to the measuring position after the calibration
process.
W30 Apply changes
button
When some settings have been changed, click on this
button in order to save the changes. Otherwise the
changes will be lost the next time the measuring head
is booted up.
W31 Time and Date
Block
Indicates the current time and date of the measuring
head. Time and date can be changed by entries into
this block.
W32 FPGA Version
Indicates the version of the FPGA.
W33 Host-Version
Indicates the version of the host system.
W34 IP-Address
Indicates the IP address of the measuring head
network adapter
W35 MAC-Address
Indicates the MAC address of the measuring head
network adapter