5

Z1

 and its accessories before cleaning (and clean only as instructed in this User Guide) 

and make sure your 

Z1

 and its accessories are dry before plugging back in.

•  Before use, make sure the power cord and plug are in good condition and that the 

cord is not damaged. 

•  Do not use bleach, chlorine, alcohol, or aromatic-based solutions, moisturizing or 

antibacterial soaps or scented oils to clean your 

Z1

Power Supply, PowerShell

 or 

Battery

. These solutions may create a shock hazard, cause damage and reduce the life 

of these items.

• 

Explosion Hazard

 — Do not use or store your 

Z1

Power Supply

PowerShell

 or 

Battery

 in the vicinity of flammable anesthetics or oxygen.

•  Do not use your 

Z1

 near toxic or harmful vapors.

•  Do not drop your 

Z1

 or any of its accessories or expose it to any crushing forces. 

Should your 

Z1

 be dropped and fail to operate properly, contact the Human Design 

Medical Service Center at 1-855-436-8729 [1-855-HDMUSA9] before using the 

Z1

 for 

treatment.

•  Do not insert any object into any tubing or opening of your 

Z1

 except as described in 

this User Guide.

•  Do not operate your 

Z1

 in any area which may block its air intake. Blocking the 

CPAP 

Tube

 and/or air inlet of your 

Z1

 while in operation could lead to overheating of the 

Z1

.

•  If you put the 

Z1

 on the floor, make sure the area is free from dust and clear of 

bedding, clothes or other objects that could block the air inlet or cover the 

Power 

Supply

 unit.

•  Only store or transport the 

Z1

 and accessories within the storage and transport 

temperature, humidity and altitude conditions specified in the Technical 
Specifications section of this User Guide.

•  Do not operate your 

Z1

Power Supply

PowerShell

 or 

Battery

 in an area which may 

cause overheating.

•  Do not cover the 

Power Supply

PowerShell

 or 

Battery

 with any object or material 

which would cause them to overheat.

•  Keep the power cord, 

Power Supply, Battery

 and 

Z1

 away from hot or flammable 

surfaces, gases or materials.

•  Do not open your 

Z1

 or tamper with its accessories. There are no user serviceable 

parts inside. Repairs and servicing should only be performed by an authorized Human 
Design Medical service agent.

•  Only power the 

Z1

 with the Human Design Medical 

Power Supply

 (AC-to-DC power 

adapter, HD60-6010) or the optional Human Design Medical 

PowerShell

 with Human 

Design Medical 

Battery

.  Do not use any other AC adapters, batteries, or other power 

sources to power the 

Z1

. Do not use the Human Design Medical 

Power Supply

PowerShell

 or 

Battery

 to power any device other than the 

Z1

.

Содержание CPAP System

Страница 1: ...Human Design Medical LLC Z1 CPAP System CONTINUOUS POSITIVE AIRWAY PRESSURE SYSTEM User Guide User Guide English...

Страница 2: ......

Страница 3: ...Box 10 Assembling and Powering Up 11 Using Your Z1 13 LCD Display 13 Starting and Stopping Treatment 14 Programming Ramp Mode 15 Inserting Removing and Replacing the Micro SD Data Card 15 User Softwa...

Страница 4: ...ng portable CPAP device Please read this User Guide carefully before using your Z1 Owner User Responsibility The owner user of the Z1 CPAP System is responsible for reading and understanding this User...

Страница 5: ...conditions Severe bullous lung disease Pneumothorax Pathologically low blood pressure Dehydration Cerebrospinal fluid leak recent cranial surgery or trauma Adverse E ects Users should report unusual c...

Страница 6: ...masks and connectors recommended by your physician clinician respiratory therapist or health care provider Your Z1 is intended to be used with CPAP masks or connectors which have vent holes to allow c...

Страница 7: ...P Tube and or air inlet of your Z1 while in operation could lead to overheating of the Z1 If you put the Z1 on the floor make sure the area is free from dust and clear of bedding clothes or other obje...

Страница 8: ...m or at least 8 feet 2 5 m above the patient It must also comply with the relevant test standard For personal computers the international standard is IEC 60950 or equivalent CAUTIONS Be careful not to...

Страница 9: ...inal and the negative terminal of the Battery to each other with any metal object such as wire Do not carry or store the Battery together with necklaces hairpins or other metal objects Do not pierce t...

Страница 10: ...se the Battery to power any device other than the Z1 Only use or charge the Battery within the operating temperature humidity and altitude conditions specified in the Technical Specifications section...

Страница 11: ...ses the Filter For proper operation of the Z1 the Filter should be replaced monthly or more often if necessary Filter The Z1 uses a specially designed filter The material has been specifically chosen...

Страница 12: ...packaging 3 Carefully inspect all contents against the Package Contents list on page 8 to assure all components are present Keep in mind that your health care provider might have completed some of the...

Страница 13: ...hrough the Filter and no air bypasses the Filter 4 Press the End Cap onto the Z1 Figure 1 NOTE To remove the End Cap squeeze firmly on the sides of the End Cap and pull it away from the Z1 Tube Adapte...

Страница 14: ...er the device is receiving power PowerShell and Battery Available Separately or as Part of Z1 Unplugged WARNING The Z1 must only be used with the optional Human Design Medical PowerShell with Human De...

Страница 15: ...within 2 3 days Using Your Z1 LCD Display Treatment Pressure Display The current treatment pressure is shown on the LCD display Once set the prescribed treatment pressure will be retained in the Z1 s...

Страница 16: ...R is displayed if there is excessive air leakage See Reminder Alert page 18 for an explanations of REMINDER conditions Starting and Stopping Treatment 1 Press the Start Stop Button on the Z1 to initia...

Страница 17: ...Replacing the Micro SD Data Card Optional Programming Ramp Mode Some users find it comforting to acclimate to CPAP therapy each session This can be accomplished by programming the Ramp Mode which all...

Страница 18: ...e Charging the Battery PowerShell Only WARNING Only charge the Battery with the supplied Human Design Medical PowerShell and the supplied Human Design Medical Power Supply Do not use any other battery...

Страница 19: ...ure they are dry before plugging back in CAUTION Do not use bleach chlorine alcohol or aromatic based solutions moisturizing or antibacterial soaps or scented oils to clean the Z1 Never attempt to cle...

Страница 20: ...e Z1 The Z1 is designed to only work with approved materials A filter which is not approved may cause damage to the Z1 from overheating CAUTION Do not operate the Z1 without the Filter The Z1 must ope...

Страница 21: ...upply is connected and the power outlet has power Loose connection of the Power Supply at the Z1 or loose connection of the Power Cord at the AC Adapter Check the connections Z1 is not locked into pos...

Страница 22: ...re not positioned correctly and mask is leaking Adjust position of the mask and headgear until leak ceases Incorrect CPAP Tube selected Do not use tubing which was not recommended by your clinician Ch...

Страница 23: ...andby CPAP and CPAP with ramp Operating pressure range 4 to 20 cm H2O Maximum single fault steady pressure state 30 cm H2O Deliverable pressure accuracies 0 6 cm H2O 4 of the measured reading Sound pr...

Страница 24: ...out by any service organization that has not been expressly authorized by Human Design Medical to perform such repairs c any damage or contamination due to cigarette pipe cigar or other smoke d damag...

Страница 25: ...provided they have been tested for Electromagnetic Compatibility EMC in accordance with the current version of RTCA DO 160 Section 21 Category M Human Design Medical has successfully completed testing...

Страница 26: ......

Страница 27: ...St Suite P Charlottesville VA 22902 About you NAME First MI Last EMAIL ADDRESS MAILING ADDRESS Street or PO Box City State Zip PHONE NUMBER About the device WHAT DEVICE DID YOU PURCHASE WHERE DID YOU...

Страница 28: ......

Страница 29: ......

Страница 30: ...Human Design Medical LLC 119 Braintree Street Suite 703 Boston MA 02134 Z1 User Guide v1 11012013 40 0007 Rev A...

Отзывы: