18
As Needed
1. Clean the exterior of the
Z1
with a soft damp cloth and mild detergent and wipe
clean. Never submerge the
Z1
in water. Do not allow soap or water to enter the
Z1
.
Filter Replacement
The
Filter
should be inspected weekly for holes and blockage by dirt and dust, and should be
replaced every month (or more often if necessary). The
Filter
is accessed by removing the
End Cap
as described in “Assembling and Powering Up:
Filter
and
End Cap
”.
NOTE:
The
Z1
uses a specially designed Filter that is only available from
Human Design Medical at 1-855-436-8724 [1-855-HDMUSA4] or online at
www.hdmusa.com.
WARNING:
Do not attempt to use any air filter which is not supplied
by Human Design Medical for the
Z1
. The
Z1
is designed to only work with
approved materials. A filter which is not approved may cause damage to
the
Z1
from overheating.
CAUTION:
Do not operate the
Z1
without the
Filter
. The
Z1
must
operate with a filter installed; failure to do so may damage the
Z1
and allow
impurities into the respiratory system.
CPAP Tube Replacement
The
CPAP Tube
should be replaced according to the instructions provided with the
CPAP Tube
.
Reminder Alert (Air Leak Detection)
“REMINDER” is displayed on the LCD screen to indicate an excessive air leak. This could be because
you are not wearing your mask. Placing the mask on your face should remove REMINDER from the
display. If you are wearing the mask and REMINDER is displayed check the following connections
for air leaks:
•
Tube Adapter
to
Z1
•
CPAP Tube
to
Tube Adapter
•
CPAP Tube
to mask
• Mask to face (adjust position of mask and headgear to eliminate the air leak)