16

Occasionally, you may be asked to remove the Micro SD Card and send it to your clinician. The 
card will contain information collected about treatment. Follow the instructions provided below 
before you attempt to remove or replace the card. 
To remove and replace the Micro SD Card: 

1. Unplug power to the 

Z1

2. To remove Micro SD Card, gently push the Micro SD Card in toward the 

Z1

 and 

release. The Micro SD Card will “pop” out from the device. Grasp the Micro SD Card 
gently and pull completely out of the 

Z1

3. Plug in the 

Z1

 

Power Supply

.

NOTE

: The 

Z1

 maintains at least 30 days of treatment history in memory 

and will update the Micro SD Card when inserted. 

User Software

See separate Z1 Software Guide for information on data viewing software. This guide may only be 
available online.

Charging the Battery (PowerShell Only)

 WARNING

: Only charge the 

Battery

 

with the supplied Human Design Medical 

PowerShell

 and the supplied Human Design 

Medical 

Power Supply.

 Do not use any other 

battery charger, AC adapter, or other device 
to charge the 

Battery

.

With both the 

Z1

 and the 

Battery

 fully inserted into the 

PowerShell

, plug the 

Power Supply

 into the 

PowerShell

 as shown 

and into a power outlet. (Figure 13)

 CAUTION:

 Make sure your 

Battery

 is fully 

charged before commencing CPAP therapy 
using the 

PowerShell

. The battery life of 

a fully charged battery in the 

PowerShell

 

is typically 8 hours or longer, after which 
treatment will terminate. To recharge the 

Battery

, see “Charging the Battery”.

Ensure the 

Z1

 is in standby mode (display reads 0.0). The 

Battery

 

will not charge unless the 

Z1

 is in standby mode.  The 

Battery

 

status indicator on the 

Z1

 will flash when the 

Battery

 is charging.  

The number of solid bars indicates the current level of charge. 
(Figure 14)

Figure 14

Figure 13

Содержание CPAP System

Страница 1: ...Human Design Medical LLC Z1 CPAP System CONTINUOUS POSITIVE AIRWAY PRESSURE SYSTEM User Guide User Guide English...

Страница 2: ......

Страница 3: ...Box 10 Assembling and Powering Up 11 Using Your Z1 13 LCD Display 13 Starting and Stopping Treatment 14 Programming Ramp Mode 15 Inserting Removing and Replacing the Micro SD Data Card 15 User Softwa...

Страница 4: ...ng portable CPAP device Please read this User Guide carefully before using your Z1 Owner User Responsibility The owner user of the Z1 CPAP System is responsible for reading and understanding this User...

Страница 5: ...conditions Severe bullous lung disease Pneumothorax Pathologically low blood pressure Dehydration Cerebrospinal fluid leak recent cranial surgery or trauma Adverse E ects Users should report unusual c...

Страница 6: ...masks and connectors recommended by your physician clinician respiratory therapist or health care provider Your Z1 is intended to be used with CPAP masks or connectors which have vent holes to allow c...

Страница 7: ...P Tube and or air inlet of your Z1 while in operation could lead to overheating of the Z1 If you put the Z1 on the floor make sure the area is free from dust and clear of bedding clothes or other obje...

Страница 8: ...m or at least 8 feet 2 5 m above the patient It must also comply with the relevant test standard For personal computers the international standard is IEC 60950 or equivalent CAUTIONS Be careful not to...

Страница 9: ...inal and the negative terminal of the Battery to each other with any metal object such as wire Do not carry or store the Battery together with necklaces hairpins or other metal objects Do not pierce t...

Страница 10: ...se the Battery to power any device other than the Z1 Only use or charge the Battery within the operating temperature humidity and altitude conditions specified in the Technical Specifications section...

Страница 11: ...ses the Filter For proper operation of the Z1 the Filter should be replaced monthly or more often if necessary Filter The Z1 uses a specially designed filter The material has been specifically chosen...

Страница 12: ...packaging 3 Carefully inspect all contents against the Package Contents list on page 8 to assure all components are present Keep in mind that your health care provider might have completed some of the...

Страница 13: ...hrough the Filter and no air bypasses the Filter 4 Press the End Cap onto the Z1 Figure 1 NOTE To remove the End Cap squeeze firmly on the sides of the End Cap and pull it away from the Z1 Tube Adapte...

Страница 14: ...er the device is receiving power PowerShell and Battery Available Separately or as Part of Z1 Unplugged WARNING The Z1 must only be used with the optional Human Design Medical PowerShell with Human De...

Страница 15: ...within 2 3 days Using Your Z1 LCD Display Treatment Pressure Display The current treatment pressure is shown on the LCD display Once set the prescribed treatment pressure will be retained in the Z1 s...

Страница 16: ...R is displayed if there is excessive air leakage See Reminder Alert page 18 for an explanations of REMINDER conditions Starting and Stopping Treatment 1 Press the Start Stop Button on the Z1 to initia...

Страница 17: ...Replacing the Micro SD Data Card Optional Programming Ramp Mode Some users find it comforting to acclimate to CPAP therapy each session This can be accomplished by programming the Ramp Mode which all...

Страница 18: ...e Charging the Battery PowerShell Only WARNING Only charge the Battery with the supplied Human Design Medical PowerShell and the supplied Human Design Medical Power Supply Do not use any other battery...

Страница 19: ...ure they are dry before plugging back in CAUTION Do not use bleach chlorine alcohol or aromatic based solutions moisturizing or antibacterial soaps or scented oils to clean the Z1 Never attempt to cle...

Страница 20: ...e Z1 The Z1 is designed to only work with approved materials A filter which is not approved may cause damage to the Z1 from overheating CAUTION Do not operate the Z1 without the Filter The Z1 must ope...

Страница 21: ...upply is connected and the power outlet has power Loose connection of the Power Supply at the Z1 or loose connection of the Power Cord at the AC Adapter Check the connections Z1 is not locked into pos...

Страница 22: ...re not positioned correctly and mask is leaking Adjust position of the mask and headgear until leak ceases Incorrect CPAP Tube selected Do not use tubing which was not recommended by your clinician Ch...

Страница 23: ...andby CPAP and CPAP with ramp Operating pressure range 4 to 20 cm H2O Maximum single fault steady pressure state 30 cm H2O Deliverable pressure accuracies 0 6 cm H2O 4 of the measured reading Sound pr...

Страница 24: ...out by any service organization that has not been expressly authorized by Human Design Medical to perform such repairs c any damage or contamination due to cigarette pipe cigar or other smoke d damag...

Страница 25: ...provided they have been tested for Electromagnetic Compatibility EMC in accordance with the current version of RTCA DO 160 Section 21 Category M Human Design Medical has successfully completed testing...

Страница 26: ......

Страница 27: ...St Suite P Charlottesville VA 22902 About you NAME First MI Last EMAIL ADDRESS MAILING ADDRESS Street or PO Box City State Zip PHONE NUMBER About the device WHAT DEVICE DID YOU PURCHASE WHERE DID YOU...

Страница 28: ......

Страница 29: ......

Страница 30: ...Human Design Medical LLC 119 Braintree Street Suite 703 Boston MA 02134 Z1 User Guide v1 11012013 40 0007 Rev A...

Отзывы: