- 30 -
When no laser is detected or you encounter some other problems with
the laser, you should do the self-test then contact and provide the error
code to Z Rapid technical support for assistance.
The system will start the self-test when the “LASER” key is switched to
the “OFF” position and the power supply is switched on. A continuous
“beeping” sound will be heard during the self-test and will cease when
the self-test is finished. The mechanical carriage returns to zero position
after the test. If no errors are found after the test, the system may be
started by switching the key to the position of “ON”.
Appendix figure 2. Self-test
It takes two to three minutes for the test. In the event the self-test is
interrupted and stops before it is completed, the laser might not be
detected when you switch the key to the “ON” position to re-start the
system. If there is no other error, this can be corrected by performing
another self-test.
Note:
If the LASER key is on the position of “ON” and the power supply is
switched on, the self-test will be skipped, and the system starts up
immediately.