iii. Using method of mesh nebulizer

1. Bottom up main unit of nebulizer, open battery cover and 
mount battery.

2. Remove medication cup module from the nebulizer.

Press PUSH key behind the main unit downwards, push medication 
cup towards main unit and remove medication cup module.

Upon removing medication cup module from nebulizer, push 
medication cup module after confirming PUSH key is pressed 
indeed, for fear of nebulizer damage.
Do  not  insert  finger  or  other  foreign  matters  from  nozzle  or 
touch spray sheet, for fear of nebulizer damage.

Attention

2).Mount battery as 
per polarity sign.

3).Close  battery 
cover.

Do not mix different types of batteries together.
When  ow  voltage  indicator  LED  (orange)  flicks,  it  means 
battery level is about to run out; at this time, replace alkaline 
battery as soon as possible.
When  ow  voltage  indicator  LED  (orange)  constantly  on,  it 
means battery level is about to run out and nebulizer fails to 
spray fog; at this time, replace alkaline battery immediately.

1). Open battery cover.

Attention

Take battery out if nebulizer is not used for a long time.
Close  the  battery  cover  immediately  after  replacing  the 
battery. Do not touch the battery cover during the atomization 
process.
Patient can contact LED indicator and battery cover within 10 
minutes.  The  maximum  temperature  on  LED  indicator  and 
battery cover is 46℃ based on maximum ambient 40℃ during 
normal operation.

Attention

Содержание M102

Страница 1: ...CE M102 180x110mm 90x110mm 2...

Страница 2: ...CO LTD Manufacturing Address Yunyang Industrial Park Danyang Jiangsu http www yuwell com TEL 86 511 86900833 Postal code 212300 European Representative Name Shanghai International Holding Corp GmbH E...

Страница 3: ...use Atomized drug used in treatment is applicable to domestic fog inhalation therapy ii Product features and indexes 1 Power supply 3V DC two AA 1 5V alkaline batteries Alternative Power adapter inpu...

Страница 4: ...using when storage temperature is below 5 The product should be stored in room with excellent ventilation and avoid violent vibration during transportation Transport should only be undertaken in the c...

Страница 5: ...imely II Using and Maintenance In order to ensure the product can be correctly used please read this Specification carefully and operate the product in strict accordance with this Specification If any...

Страница 6: ...attery cover Do not mix different types of batteries together When ow voltage indicator LED orange flicks it means battery level is about to run out at this time replace alkaline battery as soon as po...

Страница 7: ...d high viscosity suspending and volatile liquid Confirm medication cup module is well mounted it will be buckled after slight clatter sound is given off ensure electrode is excellently connected so as...

Страница 8: ...soft high frequency sound will be generated and the nebulizer will be shutdown automatically With difference of liquid feature nebulizer cannot be automati cally shut down when part of liquid is run o...

Страница 9: ...ssing through a short starting stage below 2 s Do not cover air hole with hands or other objects when mask is used in inhalation Steadily hold nebulizer with hands during inhalation There is a small a...

Страница 10: ...g after nebulization and should not be used in normal liquid inhalation 2 Cleaning 1 Main unit Wipe stains on main unit gently with wet gauze then wipe main unit with new gauze to make it dry 2 Clean...

Страница 11: ...ies used should be kept separately with main unit To avoid the strangulation keep the children away from the excessive length s Power cable If any discomfort or abnormality occurs stop using the unit...

Страница 12: ...10s Restart after medication cup module is reassembled Incline the positive surface of nebulizer to the user and make liquid contact spray sheet Clean medication cup according to operation instruction...

Страница 13: ...automatically Liquid leakage The electrode of main unit or medication cup is dirty Fault of medication cup module Power off nebulizer to clean electrode Change a new medication cup module Medication...

Страница 14: ...ectro magnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation Medical devices should also not interfere with other devices In order to r...

Страница 15: ...nment If the user of the M102 nebulizers requires continued operation during power mains interruption it is Recommendthat the M102 M103 be powered from an uninterruptible power supply or battery 3 A m...

Страница 16: ...transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with he fol...

Страница 17: ...ged caused by unforeseen natural disasters such as floods earthquakes fires etc 28 29 Table 4 Recommended separation distances Recommended separation distance between portable and mobile RF communicat...

Страница 18: ...if the temperature voltage and product characteristics are different from the indicators specified 4 Poduct performance may vary along with the liquid characteristics suspension or high viscosity 5 Se...

Отзывы: