Attention! Please
consult document
attached!
BF Applied Part
Symbol
Meaning
Meaning
Symbol
Lift up
Required to be
protected from rain
Fragile article
v. Symbols regarding safety requirement of
the machine and meaning concerned
Protection class
Ⅱ
Consider instructions
for use
Attention
Prohibit
IP22
Protected from the penetration of solid bodies with
dimensions greater than 12.5mm.protected from the
penetration of vertically tilt 15degrees falling water drops
vi. Common Fault Analysis and Troubleshooting
Fault
Cause Analysis
Troubleshooting
Too small
volume of
spray
Medication cup module
is not assembled correctly
Liquid is used up or does
not contact spray sheet
for longer than 10s
Restart after medication
cup module is reassembled
Incline the positive surface
of nebulizer to the user and
make liquid contact spray
sheet
Clean medication cup
according to operation
instructions; change a new
cup if the fault is not
removed through cleaning
Spray sheet is blocked
This item is compliant
with Medical Device
Directive 93/42/EEC
Restart after cleaning
electrode
The electrode of main unit
or medication cup is dirty
After power on,
the power
indicator LED
flashes to off
Medication cup module
is not assembled correctly
Medication cup has no
liquid or liquid is used up
Restart after medication
cup module is reassembled
Inject liquid
Incline the positive surface
of nebulizer to the user and
make liquid contact spray
sheet
Restart after cleaning
electrode
Liquid does not contact
spray sheet for longer
than 10s
The electrode of main unit
or medication cup is dirty