8. Weight: Approx. 108g (excluding batteries)
9. Size: Approx. 67 mm (L) ×48 mm (W)×125 mm (H)
10.  Safety  classification:  Do  not  use  the  Product  in  combustible 
anesthetizing  gas  mixed  with  air  or  tingled  with  oxygen  or  nitrous 
oxide.
11. Waterproof Level: IP22
12. Contraindication: It is not intended for use with Pentamidine.
13. Electrical safety requirement: Class 

 or Internally Power  Type 

BF applied part
14. Operation mode: Continuous operation
15. Normal working condition:
15. 1 Ambient temperature range: 5℃~40℃
          relative humidity range: 30%-75%
15.2 Atmospheric pressure range: 86kPa~106kPa
16.Transportation and storage environment restrictions:
16.1 Ambient temperature range: -40℃~+55℃
16.2 Relative humidity range: 10%~93%
16.3 Atmospheric pressure range: 70kPa~106kPa
 

Restore  the  equipment  to  normal      working  condition  before 
using when storage temperature is below 5℃.
The product should be stored in room with excellent ventilation 
and avoid violent vibration during transportation.
Transport  should  only  be  undertaken  in  the  condition  that 
halation devices disassemble from the main unit.
The  replacement  of  the  battery  cannot  be  operated  by  the 
patient.  It  can  only  be  operated  by  the  operator  and  cannot 
touch the patient when it is replaced. 

iii. Structure feature and working principle

With  basic  principle  shown  Fig.1.Mesh  nebulizer  is  composed  of 
medication cup and spray sheet; of which, spray sheet is spliced with 
vibrating film and circular piezoelectric component. With the circular 
piezoelectric component, electric energy is converted into mechani-
cal energy and ultrasonic vibration is generated; then vibration wave 
will  press  liquid  in  medication  cup  to  atomize  liquid  through  micro 
spray orifice on spray sheet; the fog will spray out from sprat sheet 
and onto mouthpiece or mask for inhaler use.

Attention

02

03

Содержание M102

Страница 1: ...CE M102 180x110mm 90x110mm 2...

Страница 2: ...CO LTD Manufacturing Address Yunyang Industrial Park Danyang Jiangsu http www yuwell com TEL 86 511 86900833 Postal code 212300 European Representative Name Shanghai International Holding Corp GmbH E...

Страница 3: ...use Atomized drug used in treatment is applicable to domestic fog inhalation therapy ii Product features and indexes 1 Power supply 3V DC two AA 1 5V alkaline batteries Alternative Power adapter inpu...

Страница 4: ...using when storage temperature is below 5 The product should be stored in room with excellent ventilation and avoid violent vibration during transportation Transport should only be undertaken in the c...

Страница 5: ...imely II Using and Maintenance In order to ensure the product can be correctly used please read this Specification carefully and operate the product in strict accordance with this Specification If any...

Страница 6: ...attery cover Do not mix different types of batteries together When ow voltage indicator LED orange flicks it means battery level is about to run out at this time replace alkaline battery as soon as po...

Страница 7: ...d high viscosity suspending and volatile liquid Confirm medication cup module is well mounted it will be buckled after slight clatter sound is given off ensure electrode is excellently connected so as...

Страница 8: ...soft high frequency sound will be generated and the nebulizer will be shutdown automatically With difference of liquid feature nebulizer cannot be automati cally shut down when part of liquid is run o...

Страница 9: ...ssing through a short starting stage below 2 s Do not cover air hole with hands or other objects when mask is used in inhalation Steadily hold nebulizer with hands during inhalation There is a small a...

Страница 10: ...g after nebulization and should not be used in normal liquid inhalation 2 Cleaning 1 Main unit Wipe stains on main unit gently with wet gauze then wipe main unit with new gauze to make it dry 2 Clean...

Страница 11: ...ies used should be kept separately with main unit To avoid the strangulation keep the children away from the excessive length s Power cable If any discomfort or abnormality occurs stop using the unit...

Страница 12: ...10s Restart after medication cup module is reassembled Incline the positive surface of nebulizer to the user and make liquid contact spray sheet Clean medication cup according to operation instruction...

Страница 13: ...automatically Liquid leakage The electrode of main unit or medication cup is dirty Fault of medication cup module Power off nebulizer to clean electrode Change a new medication cup module Medication...

Страница 14: ...ectro magnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation Medical devices should also not interfere with other devices In order to r...

Страница 15: ...nment If the user of the M102 nebulizers requires continued operation during power mains interruption it is Recommendthat the M102 M103 be powered from an uninterruptible power supply or battery 3 A m...

Страница 16: ...transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with he fol...

Страница 17: ...ged caused by unforeseen natural disasters such as floods earthquakes fires etc 28 29 Table 4 Recommended separation distances Recommended separation distance between portable and mobile RF communicat...

Страница 18: ...if the temperature voltage and product characteristics are different from the indicators specified 4 Poduct performance may vary along with the liquid characteristics suspension or high viscosity 5 Se...

Отзывы: