2). Open medication cup cover and then pour 
out the remaining liquid.
3). Inject a small amount (2-5 mL ) of purified 
water into medication cup and then tighten cup 
cover. Shake medication cup gently to entirely 
dissolve residual liquid in purified water.
4).Open dosing cover, pour out solution in the 
cup  and  inject  a  small  amount  (2-5  mL)  of 
purified water again.

Remove the batteriesand pull out external power adapter prior 
to cleaning and maintenance.
Since medication cup module is consumptive part, it is beyond 
our warranty. In general, service life of medication cup is half a 
year (three times or 30 min every day under normal tempera-
ture). The actual service life will vary with drug type.
Residual  liquid,  if  hard  to  be  removed,  can  be  cleaned  with 
boiled hot water or for many times.
Clean residual liquid after using every day; if not, spray sheet of 
nebulizer will be blocked and nebulization effect will be affected.
Cleaning  mode  is  only  used  in  spray  sheet  cleaning  after 
nebulization and should not be used in normal liquid inhalation.

2. Cleaning

1). Main unit: Wipe stains on main unit 
gently with wet gauze; then wipe main 
unit with new gauze to make it dry.
2). Clean electrodes on main unit and 
medication  cup  module,  so  as  to 
ensure  reliable  connection  between 
medication cup module and main unit.

Attention

14

15

5). Press startup switch continuously until indicator LEDs flicker on 
an  alternate  basis;  at  this  time,  nebulizer  will  enter  washing  mode 
and spray out purified water, so as to remove residual liquid on spray 
sheet.

Do not use volatile liquid (e.g. benzene, gasoline or diluent) to 
wipe nebulizer.
The patient can't try to touch Mesh.
Do not deliberately touch electrodes on main unit and medica-
tion cup module with cotton swab or other articles, for fear of 
electrode falling.
Remove  batteries  prior  to  parts  cleaning.  If  power  adapter  is 
used, pull off power adapter off from socket after powering off.

3). Medication cup module and 
inhalation device: Wash medication 
cup module and inhalation device 
with purified water.
4). Wipe cleaned part with new 
gauze and place the part in clean 
place to fully dry it in the air.

Attention

Содержание M102

Страница 1: ...CE M102 180x110mm 90x110mm 2...

Страница 2: ...CO LTD Manufacturing Address Yunyang Industrial Park Danyang Jiangsu http www yuwell com TEL 86 511 86900833 Postal code 212300 European Representative Name Shanghai International Holding Corp GmbH E...

Страница 3: ...use Atomized drug used in treatment is applicable to domestic fog inhalation therapy ii Product features and indexes 1 Power supply 3V DC two AA 1 5V alkaline batteries Alternative Power adapter inpu...

Страница 4: ...using when storage temperature is below 5 The product should be stored in room with excellent ventilation and avoid violent vibration during transportation Transport should only be undertaken in the c...

Страница 5: ...imely II Using and Maintenance In order to ensure the product can be correctly used please read this Specification carefully and operate the product in strict accordance with this Specification If any...

Страница 6: ...attery cover Do not mix different types of batteries together When ow voltage indicator LED orange flicks it means battery level is about to run out at this time replace alkaline battery as soon as po...

Страница 7: ...d high viscosity suspending and volatile liquid Confirm medication cup module is well mounted it will be buckled after slight clatter sound is given off ensure electrode is excellently connected so as...

Страница 8: ...soft high frequency sound will be generated and the nebulizer will be shutdown automatically With difference of liquid feature nebulizer cannot be automati cally shut down when part of liquid is run o...

Страница 9: ...ssing through a short starting stage below 2 s Do not cover air hole with hands or other objects when mask is used in inhalation Steadily hold nebulizer with hands during inhalation There is a small a...

Страница 10: ...g after nebulization and should not be used in normal liquid inhalation 2 Cleaning 1 Main unit Wipe stains on main unit gently with wet gauze then wipe main unit with new gauze to make it dry 2 Clean...

Страница 11: ...ies used should be kept separately with main unit To avoid the strangulation keep the children away from the excessive length s Power cable If any discomfort or abnormality occurs stop using the unit...

Страница 12: ...10s Restart after medication cup module is reassembled Incline the positive surface of nebulizer to the user and make liquid contact spray sheet Clean medication cup according to operation instruction...

Страница 13: ...automatically Liquid leakage The electrode of main unit or medication cup is dirty Fault of medication cup module Power off nebulizer to clean electrode Change a new medication cup module Medication...

Страница 14: ...ectro magnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation Medical devices should also not interfere with other devices In order to r...

Страница 15: ...nment If the user of the M102 nebulizers requires continued operation during power mains interruption it is Recommendthat the M102 M103 be powered from an uninterruptible power supply or battery 3 A m...

Страница 16: ...transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with he fol...

Страница 17: ...ged caused by unforeseen natural disasters such as floods earthquakes fires etc 28 29 Table 4 Recommended separation distances Recommended separation distance between portable and mobile RF communicat...

Страница 18: ...if the temperature voltage and product characteristics are different from the indicators specified 4 Poduct performance may vary along with the liquid characteristics suspension or high viscosity 5 Se...

Отзывы: