TROUBLESHOOTING

TROUBLESHOOTING

20

21

故障现象

故障现象

Concentrator is 
working, but green 
light, yellow light 
illuminate at the 
same time.

The nasal cannula 
is fogged with 
vapor or full of 
liquid drops.  
(Remove the vapor 
or liquid drops 
method: Turn on 
the device and 
connect the tubing 
to the oxygen 
outlet.  Use your 
finger to block and 
release the end of 
the tubing several 
times to remove 
the moisture and 
liquid.)

Concentrator is not 
working, the red 
light illuminate, 
buzzer alarm, the 
display show 
“LO”.

Concentrator is not 
working, the red 
light illuminate, 
buzzer alarm, the 
display show “E1”.

Concentrator is not 
working, the red 
light illuminate, 
buzzer alarm, the 
display show “E2”.

Concentrator is not 
working, the red 
light illuminate, 
buzzer alarm, the 
display show “E3”.

Solution

Solution

Probable cause

Probable cause

Symptom

Symptom

1) Clean or replace the filter.

1) Make sure the device is 
at least 10 cm away from the 
walls, other jam or heater.

1) Clean or replace the filter.

1) Clean or replace the filter.

1) Stop using the device,
 please contact the supplier
 immediately.

1) Stop using the device, 
please contact the supplier 
immediately.

1) 50% 

 Oxygen 

Concentration 

≤ 

82%.

1) Poor ventilation around 
the around the device, so 
operating temperature is 
too high.

1) Oxygen Concentration 

 50%.

1) System cycle pressure 
is too low.

1) System cycle pressure 
is too high.

1) Compressor circuit is 
opened.

Concentrator is not 
working, the red 
light illuminate, 
buzzer alarm, the 
display show “E4”.

1) Stop using the device, 
please contact the supplier 
immediately.

1) Compressor circuit is 
closed.

2) Adjust the flow rate only 
under  the advice of a 
physician.

2) Add cold water to 
humidifier.

3) Water added should 
between the maximum 
and minimum levels in 
humidifier.

4) Turn off the device 
immediately. Restart the 
device to remove the 
moisture and liquid.

5) Smooth the nasal 
cannula and tubing.

6) Replace the cooling fan.
Note: Cut off the power 
first. And    perform by 
qualified personnel only.

2) Turn off and re-turn on 
the device. Adjust the flow
 rate only under  the advice 
of a physician.

2) Oxygen flow rate over 
the maximum recommend 
flow rate: 3L/min(8F-3A,
8F-3AW); 
5L/min(8F-5A, 8F-5AW).

2) The water which added 
to humidifier is too hot.

3) Add too many water 
into humidifier.

4) The device stops 
suddenly during operating.

5) Accessory is connected 
to the device.  The tubing 
is kinked and the device 
stops suddenly.

6) Cooling fan which 
mounted inside the device 
stop running or rotate 
slowly, make the operating 
temperature is too hot.

2) Oxygen flow rate over 
the maximum recommend 
flow rate: 3L/min(8F-3A,
8F-3AW); 
5L/min(8F-5A, 8F-5AW).

3) If the phenomenon is still appeared, the device can be 
used but please contact the supplier.

7) If the phenomenon is still appeared, the device can 
be used but please contact the supplier.

3) If the phenomenon is still appeared, stop using the 
device, please contact the supplier immediately.

If the phenomenon is still appeared, stop using the 

device, please contact the supplier immediately.

Содержание 8F-3A

Страница 1: ...ang Software Park Nanjing Jiangsu Province P R China 210000 Manufacturing Address YunYang Industrial Park DanYang Jiangsu Province P R China 212300 http www yuwell com User s Manual 8F 3A 8F 3AW 8F 5A...

Страница 2: ...service to qualified service personnel Before operating the device read and understand this manual The concentrator should always be kept upright to prevent cabinet damage while being transported If t...

Страница 3: ...in the vicinity However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to other devices which can be determined...

Страница 4: ...en therapy can be hazardous Manufacturer recommends that you seek medical advice before using this product Avoid creation of any spark near medical oxygen equipment This includes sparks from static el...

Страница 5: ...fficiency 15 Safety System Over Current or Connection Loosen Unit Shut Down Compressor Over Hot Unit Shut Down Pressure Cycle Failure Alarming and Shut Down Compressor Failure Alarming and Shut Down L...

Страница 6: ...e of the concentrator is required 1 Check for any obvious damage to the carton or its contents If damage is evident please notify the carrier or local dealer 2 Remove all loose packing from the carton...

Страница 7: ...ncentrator has a damaged cord or plug if it is not working properly if it has been dropped or damaged or dropped into liquid call Qualified Service Personnel for examination and repair 2 Keep the cord...

Страница 8: ...cked or kinked tubing or a defective humidifier bottle CAUTION Meanwhile humidifier bottle will have air bubble around the sieve core Then oxygen comes from the oxygen outlet 2 Connect nasal oxygen ca...

Страница 9: ...s less than 50 Red light illuminates continuous 2 audible alarm sounds panel shows word LO and the unit shutdown Switch immediately to backup oxygen supply Call supplier immediately 4 Low High pressur...

Страница 10: ...d be disconnected before beginning preventive maintenance on the concentrator The concentrator is specifically designed to minimize routine preventive maintenance at intervals of once per year In plac...

Страница 11: ...the concentrator CAUTION TROUBLESHOOTING Power Interruption Alarm Press the power on switch on the control panel buzzer alarming display not lit device not work Concentrator is working the operating...

Страница 12: ...ot working the red light illuminate buzzer alarm the display show E4 1 Stop using the device please contact the supplier immediately 1 Compressor circuit is closed 2 Adjust the flow rate only under th...

Страница 13: ...of the device 1 Neublizer components installed incorrect 1 Do not use the remote controller in the receiving area 3 Clean or replace the nebulizer 4 Add medicinal liquid into the nebulizer according t...

Страница 14: ...mpressor Fan Breaker Exchange Valve PACKING LIST DEVICE DISPOSAL ELECTRICAL RATIONALE 1 Oxygen Concentrator 1 unit 2 User Manual 1 piece 3 Filter 1 piece 4 Nebulizer 8F 3AW 8F 5AW 1 set Contact the lo...

Отзывы: