TROUBLESHOOTING

故障现象

Solution

Probable cause

Symptom

2) Replace the nebulizer.

2) Replace the battery in 
the remote controller.

2) Neublizer deformed due 
to washing or disinfection.

2) Low battery.

*During nebulizing 
operating, the 
nebulizing
 rate is too small
 or not   working. 
(8F-3AW, 8F-5AW)

Remote control 
malfunction.

If you have all other problems, TURN OFF the concentrator 

first, use your reserve oxygen supply, and contact the 

supplier immediately.

1) Install the nebulizer 
correctly. (Figure 8).

1) Operate the remote 
controller in the front 
side of the device.

1) Neublizer components 
installed incorrect.

1) Do not use the remote 
controller in the receiving
area.

3) Clean or replace the 
nebulizer.

4) Add medicinal liquid 
into the nebulizer 
according to physician 
prescription and not over 
the maximum level.

3) Neublizer output is 
blocked.

4) Add too much medicinal 
liquid into the nebulizer.

5) If the phenomenon is still appeared, stop using 
nebulizer, please contact the supplier.

CAUTION:

OTHER ATTENTION ITEMS

Heat Exchanger

Valve Assembly

Muffler

Check Valve 

. GAS PASS OPERATION SKETCH MAP

Air

Compressor

Oxygen/Nitrogen 

Separation

Molecular 

Sieve

Oxygen

Product

Tank

Nasal Oxygen

Cannula

Nitrogen

Pressure

Regulator

Humidifier

Oxygen

Outlet 

Bacilli

Filter 

Flowmeter 

Commute Valve 

Filter

Содержание 8F-3A

Страница 1: ...ang Software Park Nanjing Jiangsu Province P R China 210000 Manufacturing Address YunYang Industrial Park DanYang Jiangsu Province P R China 212300 http www yuwell com User s Manual 8F 3A 8F 3AW 8F 5A...

Страница 2: ...service to qualified service personnel Before operating the device read and understand this manual The concentrator should always be kept upright to prevent cabinet damage while being transported If t...

Страница 3: ...in the vicinity However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to other devices which can be determined...

Страница 4: ...en therapy can be hazardous Manufacturer recommends that you seek medical advice before using this product Avoid creation of any spark near medical oxygen equipment This includes sparks from static el...

Страница 5: ...fficiency 15 Safety System Over Current or Connection Loosen Unit Shut Down Compressor Over Hot Unit Shut Down Pressure Cycle Failure Alarming and Shut Down Compressor Failure Alarming and Shut Down L...

Страница 6: ...e of the concentrator is required 1 Check for any obvious damage to the carton or its contents If damage is evident please notify the carrier or local dealer 2 Remove all loose packing from the carton...

Страница 7: ...ncentrator has a damaged cord or plug if it is not working properly if it has been dropped or damaged or dropped into liquid call Qualified Service Personnel for examination and repair 2 Keep the cord...

Страница 8: ...cked or kinked tubing or a defective humidifier bottle CAUTION Meanwhile humidifier bottle will have air bubble around the sieve core Then oxygen comes from the oxygen outlet 2 Connect nasal oxygen ca...

Страница 9: ...s less than 50 Red light illuminates continuous 2 audible alarm sounds panel shows word LO and the unit shutdown Switch immediately to backup oxygen supply Call supplier immediately 4 Low High pressur...

Страница 10: ...d be disconnected before beginning preventive maintenance on the concentrator The concentrator is specifically designed to minimize routine preventive maintenance at intervals of once per year In plac...

Страница 11: ...the concentrator CAUTION TROUBLESHOOTING Power Interruption Alarm Press the power on switch on the control panel buzzer alarming display not lit device not work Concentrator is working the operating...

Страница 12: ...ot working the red light illuminate buzzer alarm the display show E4 1 Stop using the device please contact the supplier immediately 1 Compressor circuit is closed 2 Adjust the flow rate only under th...

Страница 13: ...of the device 1 Neublizer components installed incorrect 1 Do not use the remote controller in the receiving area 3 Clean or replace the nebulizer 4 Add medicinal liquid into the nebulizer according t...

Страница 14: ...mpressor Fan Breaker Exchange Valve PACKING LIST DEVICE DISPOSAL ELECTRICAL RATIONALE 1 Oxygen Concentrator 1 unit 2 User Manual 1 piece 3 Filter 1 piece 4 Nebulizer 8F 3AW 8F 5AW 1 set Contact the lo...

Отзывы: