MAINTENANCE 

18

  

 Figure 11

  

 Figure 12

Clean Filter

1) Clean the filter with a soft cleaner or wash in warm soapy water and rinse 

thoroughly.

2) DRY the filter thoroughly before reinstallation.

3) The filter must clean or replace. 

Clean Humidifier

1) Remove the humidifier bottle from the humidifier cap then clean the bottle.

2) Remove the humidifier tube and diffuser then clean them.

 *

Clean Neublizer

Note: Must clean the neublizer after using.

1)

 

After nebulizing, remove the neublizer from oxygen concentrator.  Turn 

off  the  oxygen  concentrator,  disconnect  the  hose,  remove  the  cap, 

disassembly the neublizer as shown in Figure 8.

2)

 

Put all the nebulizer components to warm water for 15 min. (Add some 

vinegar to the warm water, if necessary.)

DO NOT

 cook or use boiling water to clean the nebulizer components.

3)

 

DRY all the components thoroughly before storage.

DO  NOT

  operate  the  concentrator  without  installing  filter, 

or  while  filter  is  wet.  These  actions  could  permanently 

damage the concentrator.

CAUTION:

TROUBLESHOOTING

故障现象

Power Interruption 
Alarm:
Press the power on 
switch on the 
control panel, 
buzzer alarming, 
display not lit, 
device not work.

Concentrator is 
working, the 
operating sound is 
normal, the 
flowmeter knob 
can be adjusted but 
no oxygen output 
or weak output.

Solution

Probable cause

Symptom

1) Re-insert the power cord 
into wall outlet securely.

1) Re-insert the power cord 
into wall outlet securely.

3) Change to another wall 
outlet.  Do not use any 
extension cord.

3) Re-install the humidifier 
tubing.

4) Press down the breaker 
reset button.

4) Replace the leaked 
accessory.

1) Power cord and wall 
outlet have poor contact.

1) Leakage between 
humidifier bottle and cap.

2) Change to another wall 
outlet.

2) Change to another wall 
outlet.

2) Wall outlet no power 
output.

2) The safety valve of 
humidifier is opened.

3) Low power of wall 
outlet.

The connection between 
oxygen output and 
humidifier is leaking.

4) Breaker reset button 
raised.

4) The accessory (nasal 
cannual, mask, humidifier, 
tubing…., etc.) is leaking.

5) If the device is still not working, please contact the 
supplier.

5) If the phenomenon is still appeared, please contact
 the supplier.

TROUBLESHOOTING GUIDE

Содержание 8F-3A

Страница 1: ...ang Software Park Nanjing Jiangsu Province P R China 210000 Manufacturing Address YunYang Industrial Park DanYang Jiangsu Province P R China 212300 http www yuwell com User s Manual 8F 3A 8F 3AW 8F 5A...

Страница 2: ...service to qualified service personnel Before operating the device read and understand this manual The concentrator should always be kept upright to prevent cabinet damage while being transported If t...

Страница 3: ...in the vicinity However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to other devices which can be determined...

Страница 4: ...en therapy can be hazardous Manufacturer recommends that you seek medical advice before using this product Avoid creation of any spark near medical oxygen equipment This includes sparks from static el...

Страница 5: ...fficiency 15 Safety System Over Current or Connection Loosen Unit Shut Down Compressor Over Hot Unit Shut Down Pressure Cycle Failure Alarming and Shut Down Compressor Failure Alarming and Shut Down L...

Страница 6: ...e of the concentrator is required 1 Check for any obvious damage to the carton or its contents If damage is evident please notify the carrier or local dealer 2 Remove all loose packing from the carton...

Страница 7: ...ncentrator has a damaged cord or plug if it is not working properly if it has been dropped or damaged or dropped into liquid call Qualified Service Personnel for examination and repair 2 Keep the cord...

Страница 8: ...cked or kinked tubing or a defective humidifier bottle CAUTION Meanwhile humidifier bottle will have air bubble around the sieve core Then oxygen comes from the oxygen outlet 2 Connect nasal oxygen ca...

Страница 9: ...s less than 50 Red light illuminates continuous 2 audible alarm sounds panel shows word LO and the unit shutdown Switch immediately to backup oxygen supply Call supplier immediately 4 Low High pressur...

Страница 10: ...d be disconnected before beginning preventive maintenance on the concentrator The concentrator is specifically designed to minimize routine preventive maintenance at intervals of once per year In plac...

Страница 11: ...the concentrator CAUTION TROUBLESHOOTING Power Interruption Alarm Press the power on switch on the control panel buzzer alarming display not lit device not work Concentrator is working the operating...

Страница 12: ...ot working the red light illuminate buzzer alarm the display show E4 1 Stop using the device please contact the supplier immediately 1 Compressor circuit is closed 2 Adjust the flow rate only under th...

Страница 13: ...of the device 1 Neublizer components installed incorrect 1 Do not use the remote controller in the receiving area 3 Clean or replace the nebulizer 4 Add medicinal liquid into the nebulizer according t...

Страница 14: ...mpressor Fan Breaker Exchange Valve PACKING LIST DEVICE DISPOSAL ELECTRICAL RATIONALE 1 Oxygen Concentrator 1 unit 2 User Manual 1 piece 3 Filter 1 piece 4 Nebulizer 8F 3AW 8F 5AW 1 set Contact the lo...